WIKA CTH63I0 Operating Instructions Manual page 25

Intrinsically safe hand-held thermometer
Table of Contents

Advertisement

Available languages

Available languages

6. Commissioning, operation
3. Use the arrow keys [▲] or [▼] to select the desired number of decimal places:
dP.
= two decimal places
dP . = one decimal place (decimal point moves one position to the right)
4. Press [ENTER/MENU] to confirm the desired setting.
Si_ appears on the display.
5. Use the arrow keys [▲] or [▼] to select the sign:
Si_
= the number to be entered is in the negative range (below 0.00 °C)
Si ┘
= the number to be entered is in the positive range
6. Press [ENTER/MENU] to confirm the desired setting.
Fd 0 appears on the display.
7. Select the range using the arrow keys [▲] or [▼]:
Fd 0 = below 1,000 °C
Fd 1 = above 1,000 °C
8. Press [ENTER/MENU] to confirm the desired setting.
00.00 appears on the display (or similar).
WIKA operating instructions, models CTH63I0 and CTH65I0
EN
25