WIKA CTH63I0 Operating Instructions Manual page 24

Intrinsically safe hand-held thermometer
Table of Contents

Advertisement

Available languages

Available languages

6. Commissioning, operation
When calling the oP1 by means of [ENTER/MENU] the oP1 function
(calibration by code) is activated even if the menu has been exited by
pressing [ESC].
EN
Example of the measured value display after entering a probe calibration code:
The small 1 in the left corner in connection with the display segment CAL in the centre of
the display indicates that the calibration by code oP1 has been activated.
Physical calibration oP2
1. Use the [▲] or [▼] arrow keys to select oP2.
2. Press [ENTER/MENU] to confirm the desired setting.
1 P is displayed in the lower display part.
3. Use the arrow keys [▲] or [▼] to choose from 1-point 1 P-, 2-point 2 P- and 3-point
3 P- calibration.
Example of 1-point calibration:
1. Confirm 1-point calibration 1 P by pressing [ENTER/MENU].
Go appears on the display.
2. Confirm by pressing [ENTER/MENU] once the measured value is stable.
After approx. 2 seconds P1 appears in the first display line for the measured value 1, dP.
appears in the 2nd line for the decimal point.
24
oP2
1P
2P
WIKA operating instructions, models CTH63I0 and CTH65I0