6. Commissioning, operation
3. Use the [▲] or [▼] arrow keys to select the channel (1 or 2) to be calibrated.
⇒
Channel selection only possible with 2-channel instruments.
4. Confirm selection by pressing [ENTER/MENU].
Unit
5. Now use the arrow keys [▲] or [▼] to select the desired calibration function.
Standard linearisation in accordance with DIN IEC 60751 "CoFF"
1. Use the arrow keys [▲] or [▼] to select CoFF.
2. Press [ENTER/MENU] to confirm the desired setting.
3. Use [ESC] to return to the measuring mode.
Calibration by code oP1
1. Use the [▲] or [▼] arrow keys to select oP1.
2. Press [ENTER/MENU] to confirm the desired setting.
⇒
A small 1 appears in the lower part of the display. 4 characters follow it (hex code /
0 ... F).
3. The 4 characters can be changed upwards by means of the [▲] arrow key.
4. Press the [▼] arrow key to go to the next position.
5. Confirm by pressing [ENTER/MENU] if all 4 characters are entered as required.
⇒
A small 2 appears and the following 4 characters can be changed as well.
6. Use [ESC] to return to the measuring mode.
WIKA operating instructions, models CTH63I0 and CTH65I0