1. Connect to the GD1 with the Ethernet or wireless connection and use a Web
browser to log into the GD1 Service Interface on
2. Enter the Diagnostics page.
3. Right Click "Download diagnostics file" and select "Save Target As..." or "Save Link
As", depending on what type of browser being used.
4. Store the file to a disc. In the file name please use the site detector ID tag number.
5. Click Done to return to main menu
6. Send the diagnostics file to SIMTRONICS support.
6.5.
Uploading a Configuration file
Simtronics can create an encrypted settings file that on upload to the GD1 can change
critical detector settings. This can be used in situations where the customer requests a
settings change, for example changing the span of the detector. The file is encrypted to
avoid non authorized persons changing critical settings. It is important to ensure that the
Configuration file belongs to the actual instrument at hand. Follow the procedure below to
upload a settings file:
Note! It is recommended to back up the initial Configuration file, by downloading it from
the instrument. If needed, this file can later be used to reset the GD1 back to the original
state. The Configuration file is downloaded on the Diagnostics page in the Service
interface.