U2701A/U2702A USB Modular Oscilloscopes User's Guide
Warm-up Procedures
Before performing any measurements, ensure that oscilloscope
has gone through the following warm- up procedures,
1 Connect the oscilloscope to the PC using a USB cable.
2 Power- up the oscilloscope.
3 Run the Agilent Measurement Manager.
4 Toggle channel 1 or channel 2 on.
5 Click on the Run button.
Test Considerations
For optimum performance, all procedures should comply with
the following recommendations,
• Ensure that the ambient temperature is stable.
• Ensure that the ambient relative humidity is less than 80%.
• Allow a 30 minutes warm- up period before calibration.
• Keep all the test connection cables as short as possible.
N O TE
Under standalone use, users are only allowed to measure up to
CAT1 30 Vrms maximum.
For high voltage measurement up to CAT1 300 Vrms, users must
install the L-mount kit on the U2701A/U2702A. Before plugging
into the instrument chassis, ensure that the L-Mount kit installed
on your modular oscilloscope is screwed to the instrument
chassis for chassis grounding purpose (see
Installation). It is required to use the provided 10:1 probes
(N2862A/N2863A) for high voltage measurements.
Getting Started
1
L-Mount Kit
3