MICRO-EPSILON thermoMETER CT Instruction Manual page 83

Hide thumbs Also See for thermoMETER CT:
Table of Contents

Advertisement

Communication Settings
12.
12.1
Baud rate:
Data bits:
Parity:
Stop bits:
Flow control:
12.2
All sensors of the CT series are using a binary protocol. They can be switched alternatively to an ASCII proto-
col. To get a fast communication the protocol has no additional overhead with CR, LR or ACK bytes.
12.3
The models CT-SF02, CT-SF15, CT-SF22, CTH-SF02, CTH-SF10 can be switched to ASCII by changing the
first figure of block 3 of the sensor calibration code. This figure has to be changed from 0 to 4 (always +4;
that means on the CTex this figure has to be changed from 1 to 5), see Chap. 7.5.
Example:
i
To switch to the ASCII protocol you can also use the following command:
Decimal:
HEX:
Data, answer:
Result:
thermoMETER CT
Communication Settings
Serial Interface
9.6 ... 115.2 kBaud (adjustable on the unit or via CompactConnect software)
8
none
1
off
Protocol
ASCII Protocol
Binary:
A6FG -
Block 1
After you have modified the sensor calibration code a reset is necessary to activate the change, see
Chap.
9.
131
0x83
byte 1
0 – Binary protocol
1 – ASCII protocol
22KB -
0AS0
Block 2
Block 3
ASCII:
A6FG -
22KB -
Block 1
Block 2
4AS0
Block 3
Page 83