Outputs and Inputs
8.3
The thermoMETER CT can optionally be equipped with a relay output. The relay board is installed the same
way as the digital interfaces, see Chap. 8.2.
A simultaneous installation of a digital interface and the relay outputs is not possible.
The switching thresholds correspondent with the values for alarm 1 and 2, see Chap. 8.5, see Chap. 8.5.2.
and are factory-set, see Chap.
Alarm 1 = 30 °C/ norm. Closed (Low-Alarm) and Alarm 2 = 100 °C/ norm. open (High-Alarm).
The adjustment of the alarms can result from the modification of the alarm 1 and alarm 2 via the program-
ming keys.
To make advanced settings (change of low- and high alarm) a digital interface (USB, RS232) and the Com-
pactConnect software is needed.