Page 5
8.2.7 Ethernet Interface ......................... 66 8.2.7.1 Installation ........................66 8.2.7.2 Installation of the Ethernet Adapter in a Network ............67 8.2.7.3 Deinstallation of an Ethernet Interface in a Network ..........68 8.2.7.4 Direct Connection to a PC ..................71 8.2.7.5 Settings inside the CompactConnect Software ............
Page 6
Emissivity ..........................95 14.1 Definition ................................ 95 14.2 Determination of Unknown Emissivity ......................95 14.3 Characteristic Emissivity ..........................96 Liability for Material Defects ....................97 Service, Repair ........................98 Decommissioning, Disposal ....................98 Appendix Optional Accessories ......................99 A 1.1 Air Purge Collar ..............................
Safety Safety System operation assumes knowledge of the operating instructions. Symbols Used The following symbols are used in these operating instructions. Indicates a hazardous situation which, if not avoided, may result in minor or moder- ate injury. Indicates a situation that may result in property damage if not avoided. Indicates a user action.
Products which carry the CE mark satisfy the requirements of the EU directives cited and the European harmonized standards (EN) listed therein. The EU Declaration of conformity is available to the responsible authorities according to the EU Directive, article 10, at: MICRO-EPSILON MESSTECHNIK GmbH & Co. KG Koenigbacher Str. 15 94496 Ortenburg / Germany The measuring system is designed for use in industrial environments and meets the requirements.
Safety Intended Use - The thermoMETER CTL is designed for use in industrial and laboratory areas. It is used for non-contact temperature measurement. - The system must only be operated within the limits specified in the technical data, see 3. - The system must be used in such a way that no persons are endangered or machines and other material goods are damaged in the event of malfunction or total failure of the sensor.
Laser Safety Laser Safety The thermoMETER CTL sensors operate with a double laser sight with a wavelength of 635 nm (visible/red), 9.4. The sensors fall within laser class 2 (II). The maximum optical power is ≤ 1 mW. Observe the laser protection regulations! Although the laser output is low, directly looking into the laser beam must be avoided.
Page 11
Laser Safety Fig. 2 Controller with laser labels If the warning label is covered over when the unit is installed, the user must ensure that supplementary label is applied. thermoMETER CTL Page 11...
Technical Data Functional Principle The sensors of the thermoMETER CTL series are non-contact measuring infrared temperature sensors. They calculate the surface temperature based on the emitted infrared energy of objects, see Chap. 12. An inte- grated double laser aiming marks the real measurement spot location and spot size at any distance on the object surface.
Technical Data Sensor Models Model Model codes Spectral response Typical applications -50 to 975 °C 8 - 14 μm Non-metallic surfaces CTLF -50 to 975 °C 8 - 14 μm Fast processes CTLM-1 485 to 2200 °C 1 μm Metals and ceramic surfaces CTLM-2 250 to 2000 °C 1.6 μm...
Technical Data Measurement Specifications 3.5.1 CTL, CTLF Models Model CTLF Temperature range (scalable) -50 ... 975 °C Spectral range 8 ... 14 μm Optical resolution 75:1 50:1 System accuracy ±1 °C or ±1 % ±1,5 °C or ±1,5 % Repeatability ±0.5 °C or ±0.5 % ±1 °C or ±1 % Temperature resolution (NETD)
Technical Data 3.5.2 CTLM-1 and CTLM-2 Models Model M-1L M-1H M-1H1 M-2L M-2H M-2H1 Temperature range (scalable) 485/1050 °C 650/1800 °C 800/2200 °C 250/800 °C 385/1600 °C 490/2000 °C Spectral range 1 μm 1.6 μm Optical resolution 150:1 300:1 150:1 300:1 System accuracy ±(0.3 % T of reading +2 °C)
Technical Data 3.5.3 CTLM-3 Models Model M-3L M-3H M-3H1 M-3H2 M-3H3 Temperature range (scalable) 50/400 °C 100/600 °C 150/1000 °C 200/1500 °C 250/1800 °C 2.3 μm Spectral range 60:1 100:1 300:1 Optical resolution System accuracy ±(0.3 % of reading +2 °C) ±(0.1 % of reading +1 °C) Repeatability Temperature resolution (digital)
Technical Data 3.5.4 CTLM-5 Model Model Temperature range 1000/2000 °C Spectral range 525 nm Optical resolution 150:1 System accuracy ±(0.3 % of reading +2 °C) Repeatability ±(0.1 % of reading +1 °C) Temperature resolution 0.2 °C Response time (90 % signal) 1 ms Emissivity/ gain 0.100…1.100...
Technical Data 3.5.5 CTLC Models Model Temperature range 200/1450 °C Spectral range 4.24 μm 3.9 μm 4.64 μm Optical resolution 45:1 System accuracy ±1 % 3 4 5 Repeatability ±0.5 % or ±0.5 °C Temperature resolution (digital) 0.1 °C Response time (90 % signal) 10 ms Emissivity/ gain 0.100…1.100...
Technical Data 3.5.6 CTLG Models Model GF-H Temperature range 100 ... 1200 °C 250 ... 1650 °C 200 ... 1450 °C Spectral range 5.0 μm Optical resolution 45:1 70:1 45:1 System accuracy ±1 °C or ±1.5 % Repeatability ±0.5 °C or ±0.5 % Temperature range (NETD) 0.1 °C Exposure time (90 % signal)
Delivery Delivery Unpacking, Included in Delivery 1 thermoMETER CTL sensor 1 Controller 1 Sensor cable 1 Mounting nut and mounting bracket (fixed) 1 Operating instructions Carefully remove the components of the measuring system from the packaging and ensure that the goods are forwarded in such a way that no damage can occur.
Optical Charts Optical Charts The following optical charts show the diameter of the measuring spot in dependence on the distance between measuring object and sensor. The spot size refers to 90 % of the radiation energy. The distance is always measured from the front edge of the sensor. The size of the measuring object and the optical resolution of the infrared thermometer determine the maximum distance between sensor and measuring object.
Page 24
Optical Charts Optics: SF D:S (Focus distance) = 75:1 16 mm @ 1200 mm D:S (Far field) = 24:1 Optics: CF1 D:S (Focus distance) = 75:1 0.9 mm @ 70 mm D:S (Far field) = 3.5:1 thermoMETER CTL Page 24...
Mechanical Installation Mechanical Installation Sensor Keep the optical path free of any obstacles. For an exact alignment of the sensor to the object activate the integrated double laser, see 9.4. The CTL is equipped with a metric M48x1.5 thread and can be installed either directly via the sensor thread or with help of the supplied mounting nut (standard) and fixed mounting bracket (standard) to a mounting device available.
Mechanical Installation Mounting Bracket The mounting bracket is included in delivery, see 4.1. 46.27 4 (.16) (1.82) (2.24) 49 (1.93) 60 (2.36) Fig. 5 Dimensions mounting bracket, fixed Dimensions in mm, not to scale The adjustable mounting bracket allows an adjustment of the sensor in two axis. thermoMETER CTL Page 39...
Electrical Installation Electrical Installation Cable Connections 7.1.1 Basic Version The basic version is supplied with a sensor cable (connection sensor - controller). For the electrical installation of the CTL please open at first the cover of the controller (4 screws). Below the display are the screw terminals for the cable connection.
Page 41
Electrical Installation Pin assignment of connector plug (connector version only) Designation Color (original sensor cable) Detector signal (+) yellow Temperature sensor brown Temperature sensor white Detector signal (-) green Ground laser (-) grey Power supply laser (+) pink not used Connector plug (outer view) Fig.
Page 42
Electrical Installation Designation (Models CTL/ CTLF/ CTLC/ CTLG) +8 ... 36 VDC Power supply Ground (0 V) of power supply Ground (0 V) of internal inputs and outputs OUT-AMB Analog output sensor temperature (mV) OUT-TC Analog output thermocouple (J or K) OUT-mV/mA Analog output object temperature (mV or mA) F1-F3...
Page 43
Electrical Installation Designation (Models CTLM) +8 ... 36 VDC Power supply Ground (0 V) of power supply Ground (0 V) of internal inputs and outputs Alarm 2 (open-collector output) OUT-TC Analog output thermocouple (J or K) OUT-mV/mA Analog output object temperature (mV or mA) F1-F3 Functional inputs Ground (0 V)
Electrical Installation Power Supply Please use a power supply unit with an output voltage of 8 - 36 VDC that provides at least 160 mA cur- rent. Residual ripple should be no more than 200 mV. Never apply voltage to the analog outputs. >...
Electrical Installation Cable Assembling The cable gland M12x1.5 allows the use of cables with a diameter of 3 to 5 mm. Remove the isolation from the cable (40 mm power supply, 50 mm signal outputs, 60 mm functional inputs). Cut the shield down to approximately 5 mm and spread the strands out. Extract about 4 mm of the wire isolation and tin the wire ends.
Electrical Installation Ground Connection 7.4.1 CTLM-5, CTLM-1, CTLM-2, CTLM-3L, CTLM-3H, CTLM-3H1 to -3H3 Models At the bottom side of the main board PCB you will find a connector (jumper), which has been placed from factory side as shown in the picture (lower and middle pin connected), see Fig.
Electrical Installation 7.4.2 CTL, CTLF, CTLC-4, CTLC-2, CTLC-6, CTLG Models At the bottom side of the main board PCB you will find a connector (jumper), which has been placed from factory side as shown in the picture (left and middle pin connected), see Fig.
Electrical Installation Exchange of the Sensor After exchanging a head the calibration code of the new sensor must be entered into the controller. After modification of the code a reset is necessary to activate the changes, see The calibration code is fixed on a label on the sensor. Do not remove this label or note the code. The code is needed if the controller must be exchanged.
Electrical Installation 7.5.1 Entering of the Calibration Code Every sensor has a specific calibration code, that is clearly printed on a label on the sensor, see Fig. 14. Fig. 14 Calibration code Please do not remove this label or make sure the code is noted anywhere. The code is needed if the sensor has to be exchanged.
Electrical Installation 7.5.2 Exchange of the Sensor Cable The sensor cable can also be exchanged if necessary. For a dismantling on the sensor side, please open at first the cover plate on the back side of the sensor. Then please remove the terminal block and loose the connections. After the new cable has been installed, please do the same steps in reverse order.
Outputs and Inputs Outputs and Inputs Analog Outputs The thermoMETER CTL has two analog output channels. 8.1.1 Output Channel 1 This output is used for the object temperature. The selection of the output signal can be done via the pro- gramming keys, see 9.
Outputs and Inputs Digital Interfaces All CTL sensors can be optionally equipped with an USB-, RS232-, RS485-, CAN Bus-, Profibus DP- or Ethernet-interface. In the case that you want to use the delivered cable gland M12x1.5 for the interface cable, please disassem- ble the terminal block and assemble them again.
Outputs and Inputs 8.2.1 USB Interface 8.2.1.1 Installation Mount the USB adapter, see 8.2. Make sure the wiring is correct according to the wire colors printed on the interface board. For industrial installations it is recommended to connect the shield of the USB adapter cable with the control- ler housing (inside the cable gland).
Outputs and Inputs 8.2.2 RS232 Interface 8.2.2.1 Installation Mount the RS232 adapter, see 8.2. Make sure the wiring is correct according to the drawing and designation printed on the interface board, see Fig. 17. The CTL always needs an external power supply for operation. 8.2.2.2 Software Installation Please install the CompactConnect software, see...
Outputs and Inputs 8.2.3 RS485 Interface 8.2.3.1 Installation Mount the RS485 adapter, see 8.2. The RS485-USB adapter is providing a 2-wire half-duplex mode. Please connect terminal A of the interface with terminal A of the next RS485 interface and so on, see Fig.
Outputs and Inputs 8.2.3.2 Sensor Installation Each CTL unit connected to the RS485 needs a different multidrop address (1 ... 32). button until M xx appears in the display. Please adjust the address by pressing the Using the Up and Down keys you can change the shown address (xx) The address can also be changed with the CompactConnect software.
Outputs and Inputs 8.2.4 Profibus Interface 8.2.4.1 Installation Mount the Profibus adapter, see 8.2. Make sure the wiring is correct, see Fig. 19. We recommend for industrial installations to connect the shield of the Profibus cable with the controller housing (inside the cable gland). The thermoMETER CTL always needs an external power supply.
Outputs and Inputs 8.2.4.2 Commissioning Profibus Read in the „IT010A90.gsd“ GSD file, contained on the delivered CompactConnect software CD, into the PLC configuration tool and configure the controller. At least one module must be selected. You will find more information about the Profibus interface on the enclosed CompactConnect software CD, page 18.
Outputs and Inputs 8.2.5 CAN BUS Interface Mount the CAN BUS adapter, see 8.2. Make sure the wiring is correct, see Fig. 19. We recommend for industrial installations to connect the shield of the CAN BUS cable with the control- ler housing (inside the cable gland).
Page 60
B4: LB B5: HB B4: DA B5: 07 T = 20.10 °C Before delivery, MICRO-EPSILON can set parameters, desired by the customer, for an extra charge. For the subsequent conversion a CAN master is required. thermoMETER CTL Page 60...
Page 61
Outputs and Inputs Diagnosis If the power supply is on, the LED displays one of the following conditions: State Meaning Flashes quickly Device is in preoperational mode. Power supply is not correct / faulty hardware. Illuminates Device is in operational mode. Sparkles Device is stopped.
Outputs and Inputs 8.2.6 Modbus RTU 8.2.6.1 Serial Interface Parameters Baud rate: 9600 or 19200, set by user (factory default: 9600) Data bits: Parity: even Stop bits: Flow control: off 8.2.6.2 Protocol The protocol is a Modbus RTU protocol. 8.2.6.3 Installation Overview Insert the Modbus RTU interface on the CTL electronic board and power it with 8 - 36 V.
Page 63
Outputs and Inputs Modbus RTU interface on CTL electronic board 8 - 36 V Fig. 23 Installation Modbus RTU interface on thermoMETER CTL electronic board thermoMETER CTL Page 63...
Page 64
Outputs and Inputs Use a Modbus RTU program to read out the data, see Fig. This is done via the Read Holding Registers and Read Input Registers, see Fig. Fig. 24 View Modbus RTU program Fig. 25 View dropdown menu Modbus RTU program thermoMETER CTL Page 64...
In order to communicate, each device needs its own ID. The numbers 1 to 247 can be selected. 8.2.6.5 Overview of Digital Commands for Modbus RTU Digital Interfaces for CT and CTLaser Sen- sors The command overview is available online on the product side of the sensor at: https://www.micro-epsilon.de/download/manuals/man--thermoMETER-ct-ctlaser-modbus-rtu-commands--en. thermoMETER CTL Page 65...
Outputs and Inputs 8.2.7 Ethernet Interface 8.2.7.1 Installation Mount the Ethernet adapter, see 8.2. In case you want to run the pre-mounted cable of the Ethernet box through the delivered cable gland, the terminal block has to be disassembled/assembled. The thermoMETER CTL always requires an external power supply of at least 12 V. Make sure the wiring is correct according to the colors printed on the interface board.
Outputs and Inputs 8.2.7.2 Installation of the Ethernet Adapter in a Network First connect the PC to the Internet. Please install the CompactConnect software CD, see If the autorun option is activated the installation wizard will start automatically. Otherwise please start CDset- up.exe from the CompactConnect software CD.
Outputs and Inputs 8.2.7.3 Deinstallation of an Ethernet Interface in a Network Select Add New Device and press Weiter. The IP and MAC address of the Ethernet adapter will appear in the list. You will find the MAC address also printed on the Ethernet adapter.
Page 69
Outputs and Inputs The following screen shows all settings. The device will be installed inside the network. Please press Fertig stellen. thermoMETER CTL Page 69...
Page 70
Outputs and Inputs To deinstall an adapter please follow the steps de- In the upper overview all on the PC installed Ether- net adapter are shown. scribed under Network Installation, see 8.2.7.2. Select Remove an Existing Device and Select the adapter(s) which should be dein- press then Weiter.
Outputs and Inputs 8.2.7.4 Direct Connection to a PC If a direct connection between Ethernet adapter and PC is required both have to be connected via a cross- over cable. In addition the adapter and the PC need to get a fixed IP address. Please open the Windows device manager after the network installation (Start/Control panel/ System/Hardware/Device manager).
Page 72
Outputs and Inputs In the input mask Use the following IP address below you can now enter a fixed IP address. thermoMETER CTL Page 72...
Page 73
Outputs and Inputs Confirm your settings with Apply. For a communication with the adapter you now have to configure the network settings on your PC. Please open the LAN settings (Start/Control panel/Network settings/Settings). thermoMETER CTL Page 73...
Page 74
Outputs and Inputs Mark the LAN connection and open the properties window using the right mouse button. thermoMETER CTL Page 74...
Page 75
Outputs and Inputs Double click on Internetprotokoll/In- Please enter here a fixed IP address for the ternet protocol (TCP/IP). Please note that the first three blocks (ex- ample: 192.168.049) have to match with the IP address of the adapter device. Press OK.
Outputs and Inputs 8.2.7.5 Settings inside the CompactConnect Software Furthermore you should set the Communication After a successful network installation of the Eth- mode Standard (menu: Measurement/ ernet adapter you can start the CompactConnect software. Settings). To make sure that an available device can be found This activates the so called polling mode (bi- you should first activate the function Scan non-...
Outputs and Inputs 8.2.7.6 Resetting the Ethernet Adapter The Ethernet adapter can be reset to the factory setting. Please use a ballpoint pen to press the reset button (hole at the top of the housing). Switch on the power supply while pressing the reset button. After a few seconds you will see a flashing green LED (network connection).
Outputs and Inputs Relays Outputs The thermoMETER CTL can optionally be equipped with a relay output. The relay board is installed the same way as the digital interfaces, see 8.2. Connect the external electrical circuit with the terminal blocks. A simultaneous installation of a digital interface and the relay outputs is not possible. The relay board provides two fully isolated switches, which have the capability to switch max 60 VDC/42 VAC RMS, 0.4 A, DC/AC.
Outputs and Inputs Functional Inputs The three functional inputs F1 - F3 can be programmed with the CompactConnect software, only. F1 (digital) Trigger (a 0 V - level on F1 resets the hold functions) F2 (analog) e = 0.1; 9 V e = 1;...
Outputs and Inputs Alarms The thermoMETER CTL has following alarm features: All alarms (alarm 1, alarm 2, output channel 1 and 2 if used as alarm output) have a fixed hysteresis of 2 K). 8.5.1 Output Channel 1 and 2 (Channel 2 on CTL, CTLG) The respective output channel has to be switched into digital mode for activation.
Outputs and Inputs 8.5.3 Open-collector Output / AL2 +24 V +24 V CT / CTlaser Relais +24 V AL 2 Fig. 28 Open-collector output / AL2 Fig. 29 Open-collector output / AL2 switching diagrams The transistor acts as switch. In case of an alarm, the contact is closed. Always one load unit (relay, LED or a resistor) must be connected.
Operation Operation After power up the unit the sensor starts an initializing routine for some seconds. During this time the display will show INIT. After this procedure the object temperature is shown in the display. The display backlight color changes according to the alarm settings, see 8.5, see 8.5.2.
Page 83
Operation Display Mode [Sample] Adjustment Range S ON Laser sighting [On] ON/ OFF 142.3C Object temperature (after signal processing) Fixed [142.3 °C] 127CH Sensor temperature [127 °C] Fixed 25CB Box temperature [25 °C] Fixed 142CA Current object temperature Fixed Signal output channel 1 [0 - 5 V] 0 - 20 = 0 - 20 mA/ 4 - 20 = 4 - 20 mA/ MV5 = 0 - 5 V/...
Page 84
Operation Display Mode [Sample] Adjustment Range ] 5.00 Upper limit signal output [0 V] According to the range of the selected output signal U °C Temperature unit [°C] °C/°F / 30.0 Lower alarm limit [30 °C] depending on model // 100.0 Upper alarm limit [100 °C] depending on model XHEAD = sensor temperature/ -40.0 ...
Operation Explanation to the Menu Items Display Description S ON Activating (ON) and Deactivating (OFF) of the Sighting Laser. By pressing the laser can be switched on and off. Selection of the output signal. By pressing the different output signals can be select- ed, see 9.1.
Page 86
Operation Fig. 31 Signal graph with P---- Red graph: TProcess with Peak Hold (Hold time = 1 s) Blue graph: TActual without post processing Display Description u 0.0 Setup of the lower limit of temperature range. The minimum difference between lower and up- per limit is 20 K.
Operation Display Description ] 5.00 Setup of the upper limit of the signal output. This setting allows an assignment of a certain sig- nal output level to the upper limit of the temperature range. The adjustment range corresponds to the selected output mode (e.g. 0 - 5 V). U °C Setup of the temperature unit [°C or °F] / 30.0...
Operation Laser Sighting The thermoMETER CTL operate with a double laser sighting which should support the sensor alignment. The measurement spot is within the two laser points. Laser radiation. In the focal point of the respective lens, see 5, both laser points overlap each other and therefore mark the Irritation or injury of the minimum measurement spot as one laser point.
Operation Error Messages The display of the thermoMETER CTL can show the following error messages: 9.5.1 CTL, CTLF, CTLC-4, CTLC-2, CTLC-6, CTLG Models - OVER Object temperature too high - UNDER Object temperature too low - ^^^CH Sensor temperature too high - vvvCH Sensor temperature too high 9.5.2...
Instructions for Operation Instructions for Operation 10.1 Cleaning Lens cleaning: Blow off loose particles using clean compressed air. The lens surface can be cleaned with a soft, humid tissue moistened with water or a lens cleaner (e.g. Purosol or B + W Lens Cleaner). Never use cleaning compounds which contain solvents (neither for the lens nor for the housing).
CompactConnect Software CompactConnect Software Insert the CompactConnect installation CD into the according drive on your computer. If the auto run option is activated the installation wizard will start automatically. Otherwise, please start CDsetup.exe from the CD-ROM. Follow the instructions of the wizard until the installation is finished. The installation wizard will place a launch icon on the desktop and in the start menu.
Communication Settings Communication Settings 12.1 Serial Interface Baud rate: 9.6 ... 115.2 kBaud (adjustable on the unit or via software) Data bits: Parity: none Stop bits: Flow control: 12.2 Protocol All sensors of the CTlaser series are using a binary protocol. Alternatively they can be switched to an ASCII protocol.
Communication Settings 12.4 Saving of Parameter Settings After power on of the CTlaser sensor the flash mode is active. It means, changed parameter settings will be saved in the internal Flash-EEPROM and will be kept also after the sensor is switched off. In case settings should be changed quite often or continuously the flash mode can be switched off by using the following command: Decimal:...
Basics of Infrared Thermometry Basics of Infrared Thermometry Depending on the temperature each object emits a certain amount of infrared radiation. A change in the temperature of the object is accompanied by a change in the intensity of the radiation. For the measurement of “thermal radiation”...
Emissivity Emissivity 14.1 Definition The intensity of infrared radiation, which is emitted by each body, depends on the temperature as well as on the radiation features of the surface material of the measuring object. The emissivity (e – Epsilon) is used as a material constant factor to describe the ability of the body to emit infrared energy.
Emissivity 14.3 Characteristic Emissivity In the case that none of the methods mentioned above help to determine the emissivity you may use the emissivity tables, see A 3, see A 4. These are only average values. The actual emissivity of a material de- pends on the following factors: - Temperature - Measuring angle...
Within this period, defective parts, except for wearing parts, will be repaired or replaced free of charge, if the device is returned to MICRO-EPSILON with shipping costs prepaid. Any damage that is caused by improper handling, the use of force or by repairs or modifications by third parties is not covered by the liability for mate- rial defects.
Service, Repair Service, Repair If the sensor, controller or the sensor cable is defec- MICRO-EPSILON MESSTECHNIK tive, please send us the affected parts for repair or GmbH & Co. KG exchange. Koenigbacher Str. 15 94496 Ortenburg / Germany In the case of faults the cause of which is not clearly identifiable, the entire measuring system Tel.
Appendix | Optional Accessories Appendix Optional Accessories All accessories can be ordered using the according part numbers in brackets [ ]. A 1.1 Air Purge Collar The lens must be kept clean at all times from dust, smoke, fumes and other contaminants in order to avoid reading errors.
Appendix | Optional Accessories A 1.2 Mounting Bracket Fig. 34 Dimensions mounting bracket, adjustable in two axes [TM-AB-CTL] Dimensions in mm, not to scale The adjustable mounting bracket allows an adjustment of the sensor in two axes. thermoMETER CTL Seite 100...
Appendix | Optional Accessories A 1.3 Water Cooled Housing Fig. 35 Dimensions water cooled housing [TM-W-CTL], hose connection: 6x8 mm, thread (fitting): G 1/8 inch Dimensions in mm, not to scale To avoid condensation on the optics an air purge collar is recommended. Water flow rate: approx.
Appendix | Optional Accessories A 1.4 High Temperature Cable For applications, where the ambient temperature can reach higher values, the usage of an optional high tem- perature cable is also recommended (ambient temperature up to 180 °C). A 1.5 Rail Mount Adapter for Controller With the rail mount adapter the CTlaser controller can be mounted easily on a DIN rail (TS35) according EN 50022.
Appendix | Factory Settings Factory Settings The devices have following presettings at time of delivery: Signal output object temperature 0 - 5 V Emissivity 0.970 (1,000 at CTLM) Transmissivity 1.000 Average time (AVG) 0.2 s (LTL); 0.1 (CTLF, CTLC-4, CTLC-2, CTLC-6, CTLG); inactive (CTLM-5, CTLM-1, CTLM-2, CTLM-3L, CTLM-3H, CTLM-3H1 to -3H3) Smart Averaging inactive (CTLF: M1, M2, M3 active)
Page 104
Appendix | Factory Settings Model M-3H M-3H1 M-3H2 M-3H3 Lower limit temperature range [°C] 1000 Upper limit temperature range [°C] 1200 1800 2000 Lower alarm limit [°C] 1200 (normally closed) Upper alarm limit [°C] 1000 1200 1600 (normally open) Lower limit signal output Upper limit signal output Temperature unit °C...
Page 105
Appendix | Factory Settings Model CTLC-2 CTLC-4 CTLC-6 Lower limit temperature range [°C] Upper limit temperature range [°C] 1450 1450 1450 1200 1650 Lower alarm limit [°C] (normally closed) Upper alarm limit [°C] 1200 1200 1200 (normally open) Lower limit signal output Upper limit signal output Temperature unit °C...
Appendix | Emissivity Table Metals Emissivity Table Metals Please note that these are only approximate values, which were taken from various sources. Material Typical Emissivity Spectral response 1.0 μm 1.6 μm 5.1 μm 8 - 14 μm Aluminum Non oxidized 0.1 - 0.2 0.02 - 0.2 0.02 - 0.2...
Appendix | Emissivity Table Non Metals Emissivity Table Non Metals Please note that these are only approximate values which were taken from various sources. Material Typical Emissivity Spectral response 1.0 μm 2.3 μm 5.1 μm 8 - 14 μm Asbest 0.95 Aphalt 0.95...
Appendix | Smart Averaging Smart Averaging The average function is generally used to smoothen the output signal. With the adjustable parameter time this function can be optimal adjusted to the respective application. One disadvantage of the average function is that fast temperature peaks which are caused by dynamic events are subjected to the same averaging time.
Need help?
Do you have a question about the thermoMETER CTL Series and is the answer not in the manual?
Questions and answers