5.3 Electronics Module Replacement

A defective electronics module can be replaced as follows:
1. Disconnect the supply power from the MT5000.
2. Remove the existing module by unscrewing the two slotted screws holding the module in the housing.
3. Note the orientation of the module and unplug from the housing base.
4. Carefully unplug the coax cable connector from side of the module.
5. Plug the coax cable into the new module.
6. Plug module onto the housing base.
7. Secure the module to the housing by screwing the two flat screws.
8. Reconnect the supply power the MT5000.
9. If a minor error in measured level is detected, the error may be compensated for using the LVL OFFSET function
in the BASIC SETUP menu. (Section 4.3.6.)
The MT5000 is now ready for use without further calibration.
MT5000-0200-1 Rev f (10-2010) DCN0528
33