Single Rod/Cable

4.3.3. PROBE LENGTH

Also described as insertion length, PROBE LENGTH is defined as the measured distance
from the first thread of the coupler (or the face of the flange) to the end of the probe. This
value must be entered in units that correspond to the UNITS of the process variable.
To set the PROBE LENGTH:
1. Press and hold the SELECT button for 2 seconds.
2. Scroll UP or DOWN to set each digit.
3. Press the SELECT button to set the digit and move to the next digit in sequence.
4. After the last digit is set, press the SELECT button to set the new PROBE LENGTH.
5. Scroll UP or DOWN to END MENU.
6. Pressing the SELECT button now will return you to the main screen.
MT5000-0200-1 Rev f (10-2010) DCN0528
Figure 9
Dual Rod/Cable
Figure 10
Probe Length
Coaxial
PROBE LENGTH
XX.XX in
LVL: XX.XX in
XX.XX
mA
12