Hach FH950 User Manual page 20

Hide thumbs Also See for FH950:
Table of Contents

Advertisement

Option
Description
Filter
Applies a data acquisition filter (Main filter or Pre-filter). The user can select the filter
parameters
parameters.
Main filter
• Fixed Period Averaging (FPA)—Fixed Period Averaging averages data over a user
• RCA time constant—The RC filter helps smooth out turbulence through the use of a
Pre-filter
• Median filter—The filtering process is done in the sensor. The feature can be disabled.
Wet/dry
Sets the sensor submersion threshold for wet and dry conditions. The default value is 20%.
threshold
Refer to
Auto zero
Sets the Auto Zero feature to On or Off.
depth
If set to On, the instrument does an air calibration when the sensor is removed from the
water and is in the air. To do the air calibration, the instrument automatically zeroes the
sensor.
If set to Off, the user can manually zero the sensor. To do this, remove the sensor from the
flow, then push and hold the Quick Jump key for five seconds.
When the sensor has been in the flow for 30 minutes, the green circle in the upper right
corner goes from green to red. This is a prompt to the user to remove and zero the sensor
again.
EMI
Sets the local line frequency for ambient noise rejection to 50 Hz or 60 Hz (default).
Clock
Sets the date and time of the portable meter in 24-hour format. Daylight savings time is not
supported.
USB
Sets the USB mode.
• Mass Storage (default)—This mode operates like a memory stick or hard drive. Files are
• CDC—This mode is used to update firmware.
Language
Selects the language used in the menus.
Units
Sets the units for velocity, flow and depth measurements. Options are Metric or English
(default).
Beeper
On (default) or Off. If set to On, the meter makes an audible tone when the sensor is at the
correct depth for applicable profile methods. The meter also makes an audible tone when
an inactive button is pushed in any menu. This feature is available only with the optional
depth sensor.
Flow
Selects the method of flow calculation for open water segment (stream profiles only).
calculation
Options are Mean-section or Mid-section. Refer to the
information.
18 English
selectable fixed period of time (1 to 480 seconds). The default is 10 seconds. If the FPA
value is 5, the velocity value shown in the display is updated once every 5 seconds.
selected time constant in the filter algorithm. This mode is useful when searching for a
maximum velocity, for example in the common 0.9 x Vmax profile method. High RC filter
time constants give higher degrees of smoothing. The time constant can be set from 2 to
20 seconds, with a default value of 6. At 1 time constant, the filter settles to
approximately 60% of the final value. At 5 time constants, the filter settles to 99.9% of
the final value. Thus, if the RC value is set to 2, the final value shows after 10 seconds.
However, the recommended (default) value is 5. Enable the feature to enter or change
this value.
Wet/Dry threshold
on page 34 for more information.
read-only.
Appendix
on page 32 for more