Footswitch jack
Coax connector
DB9 connector
Electrical line cord
Temperature probe jack
I/O Connector Description
The following I/O connections are available for remote control and monitoring. This I/O interface
provides remote monitoring and signaling to the Power Supply. Remote control can be used to
vary the Amplitude, Command Sonics Start, monitor power output and frequency.
PIN
Name
1
O.L. Indication (*1)
2
Not Connected
3
Frequency Lock (*1)
4
Frequency Output /
10
5
Power Output Watts
6
Ground
7
Sonics Start
8
Amplitude Control
9
Common Emitter
Note 1. In order to activate remote amplitude control, the Amplitude Select switch on the Back
Panel Interface PCBA must be set in the "External" position. To set the switch, power off the unit
and disconnect the power cord from the wall socket. Remove the six screws holding the cover to
the chassis. Set the switch to "External" on the Backplane Interface PCB. Re-install the cover and
fasten all 6 screws. Re-connect the power cord to the wall socket.
Please contact Qsonica for specific applications assistance related to remote control and
monitoring.
REAR PANEL
Connects to the footswitch cable.
Connects to the converter
I/O Connector
Connects to the 220/240V electrical outlet
Connects to the optional temperature probe
Description
Output - Over Load Indicator. When a power supply
overload has occurred this line will be set low referenced
to pin 9. This is an open collector output. See schematic
for ratings.
Output – Frequency Lock Indicator. When the power
supply has achieved frequency lock this line will be set
low referenced to pin 9. This is an open collector output.
See schematic for ratings.
Output – Square wave output representing the power
supply frequency divided by 10. See schematic for
ratings. This output is referenced to pin 9.
Output – Analog output representing power supply output
wattage. 5.44mV = 1 Watt. See schematic for ratings.
Ground – Common reference point for pins 2, 5, 7 and 8.
Input – Sonics Start signal. Apply a switch closure to
ground (pin 6). This will start the ultrasonic cycle as
programmed by the keypad. See schematic for ratings.
Input – Analog Input representing the Amplitude setting of
the power supply. 0 to 10 VDC signal representing 20% to
100% amplitude. See Note 1. See schematic for ratings.
Common reference point for pins 1, 3, and 4. See
schematic for ratings.