Proper setup is important to optimizing the sonication amplitude
settings and processing times. The probe should be securely
held using the Qsonica stand and clamp. This clamp can be
secured to either the booster ring or bottom of the converter.
Probe depth must be adjusted to ensure adequate mixing and
sample circulation. Larger volumes and viscous samples will
process more quickly and effectively with the addition of a mixer
or stir bar.
Probe Depth Example:
Probe Options
:
Part #
Tip Diameter
4777
1.5" (38mm)
4697
1" (25mm)
4776
0.5" (12mm)
Note: Processing volumes are application specific. There are
many variables (viscosity, concentration, etc.) that can affect the
min/max processing volumes and processing times.
In addition, the amplitude setting needed for an application must
be determined empirically.
Contact us for help with selecting the amplitude and time settings
to begin your optimization trial.
DIRECT PROBE SONICATION
Processing
Amplitude
Volume Range
100µm
5 - 20L
100µm
1 - 5L
100µm
50 - 400ml
21
It is not recommended
to insert the #4777
probe deeper than
indicated by the arrow.