Qsonica MICROSON XL 2000 Operator's Manual

Ultrasonic liquid processor

Advertisement

TM
MICROSON
X L 2 0 0 0
ULTRASONIC LIQUID PROCESSOR
53 Church Hill Road
Newtown, CT 06470 USA
203.426.0101tel
877.338.9636 toll free
203.426.7026 fax
info@sonicator.com
www.sonicator.com
1

Advertisement

Table of Contents
loading

Summary of Contents for Qsonica MICROSON XL 2000

  • Page 1 MICROSON X L 2 0 0 0 ULTRASONIC LIQUID PROCESSOR 53 Church Hill Road Newtown, CT 06470 USA 203.426.0101tel 877.338.9636 toll free 203.426.7026 fax info@sonicator.com www.sonicator.com...
  • Page 2: Table Of Contents

    T A B L E O F C O N T E N T S Section Title Page SAFETY PRECAUTIONS SAFETY REMINDER SPECIFICATIONS CONTROLS DIAGRAM INSTALLATION A. Power Requirements B. Inspection C. Placement of Equipment OPERATION A. Principle of Ultrasonics B.
  • Page 3: Safety Precautions

    S A F E T Y P R E C A U T I O N S Read ALL Instructions before Installing or Using Equipment XL-2000 Microson Ultrasonic Homogenizer Instruction Manual Your new Microson Ultrasonic Liquid Processor has been designed and tested to assure maximum operator safety.
  • Page 4: Safety Reminder

    In general use, the cable assembly should not be used to carry the convertor or pull it toward the user. Make certain the cable always has slack and is never tensioned. Move the generator or convertor assembly closer to one another to accomplish this. If this is not possible, contact Qsonica to obtain a longer assembly.
  • Page 5: Specifications

    SPECIFICATIONS Generator Dimensions: 33cm W x 19cm D x 17cm H (13" x 7.5" x 6.7") Weight: 4.2 kg (9.2 lb.) Input Voltage: 115 VAC nom. (switchable to 220 VAC nom.) 50/60 hz Full Load Current: 1.65 amps at 115v/ 0.8 amps at 220v Fuse Rating and Type: 3 amp/1.5 amp Fast acting Voltage Tolerance: +/- 10% nominal voltage Output Voltage/Frequency: 950 Vrms (maximum) 22.5 kc (nom)
  • Page 7: Installation

    I N S T A L L A T I O N WARNING! High voltage is present in the generator, convertor and high frequency cable. Do not attempt to remove the generator cover or convertor case. Do Not touch any open cable connections on the unit while the power is turned on. Power Requirements The generator requires a single phase, grounded, three-wire, 115V or 220V, 50/60 Hz source, unless otherwise fitted, and has a 3 amp fuse for 115V service or a 1.5 amp fuse for 220V...
  • Page 8: Inspection

    Unpack the unit from its shipping carton and check the contents against the packing list. Before disposing of the packing material, check it carefully for small items. Report any missing components to Qsonica immediately. Visually inspect all external controls, indicators, and surfaces to detect any damage in transit. If damage has occurred, contact your carrier within 48 hours of delivery date.
  • Page 9: Operation

    O P E R A T I O N Principle of Ultrasonics The generator (power supply) converts conventional 50/60 Hz AC line power to 22.5 kHz electrical energy which is fed to the convertor where it is transformed to mechanical vibration.
  • Page 10: Preparation For Use

    Preparation for Use The Microson is easy to set-up. Simply attach one end of the convertor cable to the back of the generator. Plug one end of the power cord into the back of the unit and the other end into a three-pronged (grounded) wall outlet. Then make sure that the MICROPROBE is securely attached to the convertor.
  • Page 11 Pulsed Operation Press the Remote operation switch on the front panel Pulse as desired using transducer pulse button switch with appropriate intensity power setting Footswitch Operation Switch front panel switch to Continuous operation Insert footswitch plug into jack on rear of the unit Set desired power intensity level Pulse as desired using footswitch The sides and end of the tip must never be allowed to come in direct contact with the...
  • Page 12: Care Of Microprobe Tips

    Insert Timer plug into the footswitch socket on rear panel of generator. Adjust the power setting to the appropriate intensity level for your sample. Set the desired time on the timer. Once timer is set the probe will be activated. Unit will shut off at end of timed cycle.
  • Page 13: Repair And Replacement

    MICROTIP probes and pack separately in same outer carton. IMPORTANT NOTICE By returning any material to Qsonica, the customer or the customer's agent thus certifies that any and all materials so returned are, or have been rendered, free of any hazardous or noxious matter or radioactive contamination and are safe for handling under normal repair shop conditions.