Alam Mode; Alam - Optex THERMO-HUNTER BUILT-IN Instruction Manual

Non-contact thermometer
Table of Contents

Advertisement

Setup of Functions

ALAM

ALAM/ Alarm (contact) output mode
ALAM
400
OUT
NOR
NOR
OUT
DEL
DEL
OUT
ONES
ONES
16
You can set the temperature and output method of
ALAM
alarm (contact)output.
H (upper limit value): Output turns on when the value
50
exceeds the Setting value.
L (lower limit value): Output turns on when the value
falls below the Setting value.
Select one of the three types of OUT (output) mode:
NOR, DEL and ONES.
NOR: Output is kept on as long as the value is over or below
the set temperature.
Contact output
DEL: Output turns on when a specifi c time period passes
after the value exceeds or falls below the set temperature.
Contact output
ONES: Output turns on when the value exceeds or falls
below the set temperature and is kept for a specifi c time.
Contact output
H
L
ON
OFF
H
L
ON
OFF
t
t: Time to delay output
H
L
ON
OFF
t
t: Time to keep output on
Measurement value
Measurement value
Measurement value