Setup of Functions

BANK

BANK/ Bank mode
BANK
1
BANK
4

DELY

DELY/ Delay mode
DELY
1
DELY
200

TADJ

TADJ/ Thermo Adjust mode
TADJ
E1. 0 0
Indications of selectable settings
TECH
E1. 0 0
TECH
12
Select the bank No. to make the setting.
There are four banks (1 to 4) in total, each of which can have its own
setting.
The No. displayed fi rst is that of the bank enabled in normal
operation.
* The bank enabled in normal operation should be selected by
connecting the cable.
You cannot select it in the setting mode.
(1) Enter the BANK mode select the bank No.
(2) Press the ENTER button to confi rm the setting.
* Switch banks in END mode, when changing the setting value for
each bank in succession.
* The changed values for setting become effective by saving them in
END mode(SET).
Select the response time.
You can select the value between 1 and 200 of the number of moving
average times. Selecting a larger value will delay the response time
more.
1 = response time of the product (0.15 sec.) to 200 = approximately
10 sec.
This setting can average (smooth) fl uctuation of measurement
values and large variation of temperature.
(1) Enter the DELY mode select the value.
(2) Press the ENTER button to confi rm the setting.
Make the temperature measurement settings.
TECH: Input the temperature value of the target to automatically
calculate the emissivity rate.
ε
:
Input the emissivity rate directly.
AADJ: The display value can be adjusted in accordance with the
specifi ed value (within the measurement temperature range).
NON: Cancel the TADJ mode. The unit returns to the function
selection mode.
E
E1. 0 0
ε
AADJ
NON
OFF
SET
AADJ
* Setting NON with SET returns the unit to
the state of before making the setting.
NON