Setup of Functions
AADJ
AADJ/ Analog Adjust mode
AADJ
OFF
AADJ
ON
EXIt
EXIT
EXIt
* If the timings to set L and H are different (e.g., when using the same target for setting L and
H), set either of them fi rst and save the setting with END. Otherwise the input value will be
canceled.
300
AADJ
14
This setting is for adjusting the measurement value with the specifi ed value
according to the measuring targets. A intended value can be output by setting
L (lower limit value) and H (upper limit value) conforming to the both specifi ed
value.
* Measurement accuracy can not be guaranteed for the value made by AADJ
mode.
* Please do not change emissivity after setting in the AADJ mode, otherwise
the adjusted value will be changed.
L
SET
(1) Set the AADJ mode to ON and press the ENTER button to make
the setting.
L
(2) Press the ENTER button after confi rming the target is aimed
suffi ciently larger than the fi eld of view.
(3) Check that the target is in the measurement area. The current
measurement value appears on the display. Press the ▲/▼
buttons to input the temperature L (lower limit value) of the target.
(4) Press the ENTER button to confi rm the setting.
(5) Check that the indicated value has been changed.
* To input the value later, select EXIT.
H
SET
(1) Set the AADJ mode to ON and press the ENTER button to make
the setting.
(2) Press the ENTER button after confi rming the target is aimed
H
suffi ciently larger than the fi eld of view.
(3) Check that the target is in the measurement area. The current
measurement value appears on the display. Press the ▲/▼
buttons to input the temperature H (upper limit value) of the target.
(4) Press the ENTER button to confi rm the setting.
(5) Check that the indicated value has been changed.
* To input the value later, select EXIT.
The adjusted value are effective after the both of L and H value are set
and stored. When AADJ mode on, AADJ is displayed as sub indication.
The measuring value is output according to the setting value previously
stored, if the setting of L and H value are stored in EXIT mode with AADJ
mode on. In this process, AADJ is also displayed as sub indication.
* Either the upper or the lower limit value can not be changed after the
setting is completed.
120
100
120
SET
250
300
300
SET