• Visually inspect for damage and/or wear.
• Check the action of moving parts to ensure smooth operation throughout the intended range of motion.
Note: If damage or wear is noted that may compromise the function of the instrument, do not use.
• Steam sterilize only. Follow instructions in STERILIZATION RECOMMENDATIONS.

CLEANING INSTRUCTIONS FOR THE POWER SUPPLY

• Ensure the power supply has been fully disconnected from the power source.
• Clean the power supply with a damp, lint-free cloth using a pH neutral detergent.
• An alcohol wipe may be used to disinfect the surface of the power supply.
• Never immerse the power supply.

STERILIZATION RECOMMENDATIONS

Steam sterilization is safe and effective, and is the only recommended method for the dermatome. There
are no contraindications for sterilizing the Zimmer Electric Dermatome and accessories.
Ethylene oxide sterilization is not recommended because reliable outgassing times are difficult to
determine for lubricated powered instruments.
Never sterilize the power supply. Disconnect the dermatome from the power supply before sterilization.
Place cleaned instruments in an instrument tray or fully perforated autoclave case. The Zimmer
Dermatome Autoclave Case (REF 00-8801-003-00) is recommended. (See Fig. 18.)
Do not kink or crimp the dermatomeʼs cord when closing the case lid.
If the Zimmer Dermatome Autoclave Case is used, the instruments are to be wrapped, two
double thicknesses of #140 thread count wrappers, or equivalent. If sterilization wraps are used,
they must be free of detergent residues. Foam sheets should not be reused. They may have
trapped impurities from the steam supply, and subsequently may form deposits on the
instruments. Textiles that have been scorched by overheating also may form deposits on
instruments. Exposure times are the same for wrapped or unwrapped instruments.
Follow instructions in RECOMMENDED STEAM STERILIZATION PARAMETERS.
Do not immerse in liquid to cool. Cool by exposure to room temperature or cover with a cold,
sterile towel.