The user and operating staff must always pay close attention to the CLEANING PRECAUTIONS and
CLEANING INSTRUCTIONS FOR THE DERMATOME. Failure to follow these instructions may damage
the dermatome.

IMPORTANCE OF THE NEED TO ADHERE TO A CARE REGIMEN

The handpiece and accessories must be inspected prior to each use.
• Visually inspect for damage and/or wear.
• Always inspect the handpiece carefully for possible scratches, nicks, or burrs caused by extended use
or mishandling.
• Inspect the dermatomeʼs cord for cuts or missing insulation caused by extended use or mishandling.
• Check the action of moving parts to ensure smooth operation throughout the intended range of motion.
• Annual calibration checks are strongly recommended to verify continued accuracy. Note: If damage or
wear is noted that may compromise the function of the instrument, do not use.

SETUP INSTRUCTIONS

• Observe sterile field precautions per hospital protocol.
• Connect the Zimmer Electric Dermatome to the power supply by inserting the connector plug into the
connector receptacle. (See Fig. 7.) Remove any kinks or twists from the cord. Align the connector plug to
the receptacle by rotating the plug while gently pushing inward. When aligned, push in until the plug clicks
into the receptacle.
Note: Older Zimmer Dermatomes and Power Supplies were equipped with a metal connector plug and
receptacle. The metal connectors have been upgraded to plastic connectors for ease of use. The older
metal connectors are not compatible with the plastic connectors.
• Before connecting the power supply to a power source, be sure that the switch on the front of the power
supply is in the OFF (O) position and that the safety lock on the Zimmer Electric Dermatome is in the
SAFE position. Connect to a power source via the "IEC" connector in the rear of the power supply.
• Turn power on by moving the rocker switch on the front of the power supply to the ON (I) position. The
rocker switch should illuminate. Activate the dermatome by completely depressing the throttle lever with
the safety lock in the ON position. Return the safety lock to the SAFE position, when not in use.
• During set-up procedure, visually inspect for damage and/or wear. If damage or wear is noted that may
compromise the function of the instrument, do not use.
BLADE INSTALLATION (See Fig. 5.)
• Use a new sterile blade for each procedure. Use only Zimmer Dermatome Blades (REF 00-8800-000-
10).
• To install blade: