ENGLISH (US-VERSION) CONTENTS page I I m m p p o o r r t t a a n n t t I I n n f f o o r r m m a a t t i i o o n n General _____________________________________________________________4 Safety ______________________________________________________________4 Transport ___________________________________________________________4...
Important Information G G e e n n e e r r a a l l • Ultra Deep Freezers of the UF-series are designed • Keep these operating instructions ready at hand to ensure the structure-maintaining long-term and leave them with the machine, so that all storage of valuable preparations at very low users can find out about the functions and safety temperatures.
Installation min 70 mm min 70 mm L L o o c c a a t t i i o o n n POWER FAIL ALARM HISTORY POWER FAIL ALARM HISTORY ENTER ENTER POWER FAIL ALARM HISTORY ALARM TEST ALARM TEST ENTER ALARM TEST •...
S S w w i i t t c c h h i i n n g g o o n n • Check the setting of the circuit breaker. (see • After the self-test, the temperature inside the chapter “ Fuse / Over-current Release ”) refrigerator is displayed.
Page 8
DATE/TIME SET DATE/TIME HH:MM dd MMM yy Setting current time and date. SELECT 12h/24H 12h DISPLAY 24h DISPLAY Changing between 12-hour and 24-hour display ALARMS SET LOW ALARM LOW ALARM : X.X °C The lowest admissible internal temperature. If the temperature falls below this value a temperature alarm is triggered.
USER PASSWORD Changing the password to access the “SETTINGS” MENU. CHANGE USER PASSWORD: 0XXX XXXX Enter the new password twice and confirm. CALIBRATION This function is for calibrating the sensors. As each setting in the calibration compromises the function of the appliance, this function is protected by a password in the factory and must only be used by trained service staff!
VOLTAGE IND. LINE VOLTAGE NONE Shows the input voltage of the appliance on the display. AMBIENT PROBE AMBIENT NONE If an ambient temperature sensor is connected to the appliance , the temperature measured by this sensor can be indicated on the display. ADD.PROBE Enables the additional display of the temperature values of one of the following sensors (if connected):...
ALARM TEST Appears during an alarm test. The message disappears as soon as the test has finished. AMBIENT ..°C If an ambient temperature sensor is connected to the appliance and the display in the EXTRA menu activated ( - EXTRA - AMBIENT PROBE), this message appears followed by the currently measured ambient temperature in the display.
BAT.1 LOW/FAIL This message appears if the internal main battery is faulty or flat. At the same time the battery symbol also appears on the display. BAT.2 LOW/FAIL This message appears if the optional additional battery is faulty or flat. At the same time the battery symbol also appears on the display.
A A l l a a r r m m H H i i s s t t o o r r y y • The alarm history list contains all relevant data • The alarm history list can be accessed via the on temperature and powerfail events.