HP 8712ET User Manual Supplement page 48

Automating measurements, rf network analyzers
Hide thumbs Also See for HP 8712ET:
Table of Contents

Advertisement

Operator Interaction
Using a Foot or Hand Switch
210
ELSE
220
ASSIGN @Hp8711 TO 716
230
ABORT 7
240
CLEAR 716
250
END IF
260
!
270
Pass_count=0
280
Start: !
290
LOOP
300
! Display message
310
Msg$="'DUTs passed: "&VAL$(Pass_count)&CHR$(10)
320
Msg$=Msg$&"Press button to measure next DUT.'"
330
OUTPUT @Hp8711;"DISP:ANN:MESS ";Msg$
340
!
350
! Wait for button to be pressed
360
REPEAT
370
OUTPUT @Hp8711;"DIAG:PORT:READ? 15,1"
380
ENTER @Hp8711;X
390
UNTIL X = 0
400
DISP "Button is now pressed."
410
OUTPUT @Hp8711;"DISP:ANN:MESS:CLEAR"
420
!
430
! Wait for button to be released
440
REPEAT
450
OUTPUT @Hp8711;"DIAG:PORT:READ? 15,1"
460
ENTER @Hp8711;X
470
UNTIL X = 1
480
DISP "Button is now released."
490
!
500
OUTPUT @Hp8711;"DISP:ANN:MESS 'Measuring...'"
510
! Add code here to take sweep
520
! and measure DUT.
530
WAIT 1
540
Pass_count=Pass_count+1
550
END LOOP
560
END
3-20
Automating Measurements

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8712esHp 8714etHp 8714es

Table of Contents