Example For Agilent Technologies 81650A; Output Power And Stability Test; Output Power Test - Agilent Technologies 81650A User Manual

Laser source modules
Table of Contents

Advertisement

Performance Tests
3DUDPHWHU
Mean wavelength
FWHM
36

Example for Agilent Technologies 81650A

0LQ 6SHF
1295.0 nm

Output Power and Stability Test

Mainframe 8163A w/ 8165xA (DUT)
and 81533B + 81525A (Ref)
8165xA 81533B
81113PC
Connector
slanted
Figure 7
Measurement Setup for Output Power and Stability Test

Output Power Test

1 Make absolutly sure that cable connectors and detector windows
are clean.
2 Enable the laser output of the Agilent 8165xA and allow
instruments to warm up for at least 20 minutes.
3 On the power meter:
a Set the wavelength of the power sensor exactly to the wavelength of the
laser source.
b Ensure that the [C
c
With the connector adapter on the optical head, covered with a plastic
cap, zero the powermeter.
d Set the averaging time to 500 ms.
4 Connect the equipment as shown in figure D-2.
5 On the Agilent 8163A with the module under test (DUT):
Agilent 81650A/1A/2A/4A/5A/6A/7A User's Guide, Third Edition
5HVXOW
1309,46 nm
2,87 nm
8163A
Optical Head 81525A
Ref
] parameter on the powermeter is set to zero.
AL
Performance Test
0D[ 6SHF
1325.0 nm
< 3.5 nm