Overview KLE 325 Overview Structure and function Structure Measuring electrode Temperature sensor in graphite enclosure Shaft Closing head Cable screw joint Recommended fields of application On site measurements in rivers, lakes and wastewater Fish farming Ground water measurements Applications in water laboratories...
KLE 325 Cleaning Cleaning CAUTION To clean the sensor, disconnect it from the instrument. Outside cleaning We recommend to clean the sensor thoroughly, especially before measuring low conductivity values. Contamination Cleaning procedure Lime sediments Immerse in acetic acid for 5 minutes (volume...
Sensor with connection IP 68 (2 x 10 Pa or 2 bar) resistance cable Cable plug IP 67 (when plugged in) The KLE 325 meets the requirements according to article 3(3) of the directive, 97/23/EC ("pressure equipment directive"). ba75798e02 01/2009...
Page 7
KLE 325 Technical data Measurement Conductivity measuring 10 S/cm ... 20 mS/cm conditions range Temperature range 0 ... 80 °C (32 ... 176 °F) Max. allowed overpressure 2 x 10 Pa (2 bar) Minimum depth of 36 mm immersion Maximum depth of...
Page 8
Technical data KLE 325 ba75798e02 01/2009...
Need help?
Do you have a question about the KLE 325 and is the answer not in the manual?