10. Connect the pyrolyzer to the detector with
the three provided mounting screws (see
Diagram 9-6).
Diagram 9-6. Attaching the pyrolyzer.
terminal
module
socket
(con5)
connector
tubes
power switch
rectangular access at
bottom of bracket
mounting
screws (3)
pyrolyzer
Midas
WARNING
when using the pyrolyzer, perform a gas calibration
every six months. Do not allow the ambient tem-
perature at the point of installation to exceed 40°C
(104°F). Operation above this temperature may require
more frequent bump testing or calibration to confirm
working specification. Because of the higher operat-
ing temperatures when using the pyrolyzer module,
Honeywell Analytics strongly recommends that the
ventilated Midas Top Cover (shown in Diagram 9-1) be
used in all pyrolyzer applications.
Note:
Consult the label on the side of the Pyrolyzer
module, as shown in Diagram 9-7, for the
firmware version level required in the Midas
detector for proper operation.
Diagram 9-7. Pyrolyzer label.
P/N: MIDAS-T-NP1
05/16 REV 1 S/N:00001
For non-flammable gases only
www.honeywellanalytics.com
This pyrolyzer is to be used only
with Midas extractive units
equipped with version 1.10 or later
application software
9-5
Gas Detector
®
To maintain sensor accuracy
.