Chapter 10 – Malfunctions
209
Process was canceled
manually
301
Overflow switch active
302
Fresh water supplied even
though inlet valve is closed
303
Overheating
304
Temperature too low
305
Circulation pump defective
306
Temperature sensor -
Heating element
307
Temperature sensor -
Chamber
308
Internal error
Cleaning and disinfection process was
stopped by user
Unit is leaking (e.g. water, wash,
processing media)
Inlet valve is defective
Electronics are defective
Heating element defective
Foreign bodies in pump wheel
No water
Temperature sensor defective
Temperature sensor defective
Software error
Missing component
Cleaning and disinfection process
was not completed!
Program must be restarted
Turn off water supply!
Be sure to inform specialist
supplier/customer service!
Fluids in the unit, e.g. in the base pan,
may contain processing media in case
of damage. Remember to follow the
safety precautions of the processing
Immediately turn off water supply!
media manufacturer
Contact customer service
Contact customer service
Check water supply
Contact customer service
Contact customer service
Turn off unit and restart
Call customer service