Chapter 10 – Malfunctions
General situation
Remaining moisture on
instruments
Remaining moisture in
hollow instruments
Possible causes
Base rack and/or basket inserts
or racks not loaded properly or too
full
Door was closed for a long time after
program ended (see logs)
Recommended course of action
Arrange washware properly and do
not overload
Dry instruments manually
Open the door immediately after program ends.
Dry narrow-necked instruments manually
with sterile compressed air