Chapter 10 – Malfunctions
General situation
Knocking or clattering
noises in the washing
chamber
White deposits on the
instruments
Spots on the
instruments
Unsatisfactory
cleaning results
Unit does not work;
there is nothing on the
display
56
Before contacting us
Follow the steps as indicated for dealing with a warning or a malfunction
message appearing on the display of the HD 450/HD 450 Injection. In
addition, you will find
the most important occurrences in the following table. These situations
are listed next to possible causes and the corresponding courses of
action.
Should you not find your situation in the table below or should the
actions you take be unsuccessful, please contact an authorized
representative. To better help you, please have the serial number of
your HD 450/HD 450 Injection ready, as well as a detailed description
of the malfunction.
Possible causes
Washing arm is hitting the
instruments and vessels
The water softening system
has not been programmed
The deposits on the instruments
are from the remains of dental
cement
Neutralization agent is missing
Base rack and/or basket inserts
or racks not loaded properly or too
full
Arrangement leads to spray shadows
Dirt and stains are too dry
Verschmutzungen
Washing arm nozzles or nozzles
of injector rail are clogged
Screens in the pump
sump are dirty
Unit is not turned on Power plug
not plugged in Circuit breakers
have tripped
Recommended course of action
Cancel program, rearrange washware
and restart program
Check the hardness of the tap water and if
necessary, reprogram the water softening
system
Manually clean the instruments before the
cement hardens
Add rinsing agent
Arrange washware properly
according to recommended load and
do not overload
Avoid spray shadows
Do not allow dirt and stains to dry for more
than 6 hours
Unclog nozzles
Clean the wide- and fine-mesh screens
Turn on unit at the power switch
Plug in the power plug
Reset the breakers (see type label for
min. fuse protection)