Adjustment screen with Indicating Tools

If you are using an indicating tool, or if you use the peak mode button to change modes manually, the
adjustment screen will look like this:
In this case, the last cycle results are disabled (as indicating tools do not mark attainment of target torque in
a way that the TWC can detect, such as by clicking) and the cycle limit settings are disabled (as the TWC
cannot automatically cycle indicating tools since it cannot detect when the target is reached).
The adjustment screen works exactly the same as it does for setting tools, except for the cycle and
transducer dial; here, both the single and multiple cycle buttons initiate an auto-load sequence to 80% of the
currently set target torque. Because multiple cycles are not supported for indicating tools, the multi-cycle
button behaves the same as the single-cycle button.