Standard Definitions Version Changes - Agilent Technologies 85038M User's And Service Manual

7-16 calibration kits
Table of Contents

Advertisement

Standard Definitions
Version Changes
Version Changes
Class assignments and standard definitions may change as more accurate model and
calibration methods are developed. The disk shipped with the kit will contain the most
recent version.
Standard Class Assignments
Class assignment organizes calibration standards into a format compatible with the error
models used in the measurement calibration. A class or group of classes corresponds to the
systematic errors to be removed from the measured network analyzer response.
and
Table A-2
list the classes of the devices in these calibration kits. This information
resides on the calibration definitions disks included in these kits.
Table A-1
Standard Class Assignments for 8753 and 872x Families
Calibration Kit Label: 7-16 or 85038
Class
S
A
11
S
B
11
S
C
11
S
A
22
S
B
22
S
C
22
Forward transmission
Reverse transmission
Forward match
Reverse match
Response
Response and isolation
A-2
A
B
C
D
E
2
1
3
2
1
3
4
4
4
4
1
2
4
1
2
4
F
G
Standard Class Label
Opens
Shorts
Load
Opens
Shorts
Load
Thru
Thru
Thru
Thru
Response
Response & Isol'n
Table A-1
85038A/F/M

Advertisement

Table of Contents
loading

This manual is also suitable for:

85038a85038 series85038f

Table of Contents