Specifications Environmental Requirements; Temperature-What To Watch Out For - Agilent Technologies 85038M User's And Service Manual

7-16 calibration kits
Table of Contents

Advertisement

Specifications
Environmental Requirements
Environmental Requirements
Table 2-1
Environmental Requirements
Parameter
a
Operating temperature
Error-corrected temperature range
Storage temperature
Altitude
Operation
Storage
Humidity
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the
range over which the network analyzer maintains its specified performance while correction
is turned on.
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range shown in
Table
2-1.
IMPORTANT
Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
2-2
Limits
+20 °C to +26 °C (+68 °F to +79 °F)
b
±1 °C of measurement calibration temperature
0 °C to +55 °C (−32 °F to +131 °F)
< 4,600 meters (≈15,000 feet)
< 15,300 meters (≈50,000 feet)
90% for 24 hours
85038A/F/M

Advertisement

Table of Contents
loading

This manual is also suitable for:

85038a85038 series85038f

Table of Contents