Troubleshooting Troubleshooting Process - Agilent Technologies 85038M User's And Service Manual

7-16 calibration kits
Table of Contents

Advertisement

Troubleshooting
Troubleshooting Process
Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance
verification, follow the steps in
Figure 5-1
Troubleshooting Flowchart
Everything
is Okay
pn43m
5-2
Figure
5-1.
Start
(Bad Cal Suspected)
NO
Bad
Measurement?
YES
NO
Devices
Cleaned?
YES
Proper
NO
Torque?
YES
Correct
NO
Cal Constants
Loaded?
YES
Mating
YES
Surfaces Smooth &
Unmarred?
NO
Replace Device(s)
Clean Devices
Torque 7-16 Devices
to 20 in-lb, 226 N-cm
Load Correct Cal Constants
Recertify Kit
Recalibrate System
85038A/F/M