Recertification; How Often To Recertify; Where To Send A Kit For Recertification - Agilent Technologies 85038M User's And Service Manual

7-16 calibration kits
Table of Contents

Advertisement

Recertification

The following will be provided with a recertified kit:
• a new calibration sticker affixed to the case
• a certificate of calibration
• a calibration report for each device in the kit listing measured values, specifications,
and uncertainties
NOTE
A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
Agilent Technologies offers a Standard calibration for the recertification of your kit. For
more information, contact Agilent Technologies. For contact information, see
page
5-3.

How Often to Recertify

The suggested initial interval for recertification is 12 months or sooner. The actual need for
recertification depends on the use of the kit. After reviewing the results of the initial
recertification, you may establish a different recertification interval that reflects the usage
and wear of the kit.
NOTE
The recertification interval should begin on the date the kit is first used after
the recertification date.

Where to Send a Kit for Recertification

Contact Agilent Technologies for information on where to send your kit for recertification.
For contact information, see
When you return the kit, complete and attach a service tag. Refer to
Device to Agilent," on page 5-3
85038A/F/M
Table 5-1 on page
5-3.
for details.
Performance Verification
Recertification
Table 5-1 on
"Returning a Kit or
4-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

85038a85038 series85038f

Table of Contents