Omega DP400TP User Manual page 35

Strain indicator
Hide thumbs Also See for DP400TP:
Table of Contents

Advertisement

5
Full-scale
Full-scale value of strain-gauge sensor, used during calibration type
"Calibration on full-scale % value".
-999999 +999999 [digit
6
Lower limit
Lower limit of sensor, used during input calibration type Sampling value
or Calibration on full-scale % value.
-999999 +999999 [digit
7
Upper limit
Upper limit of sensor, used during input calibration type Sampling value
or Calibration on full-scale % value
-999999 +999999 [digit
8
Offset calibration
Analog input offset correction. Value added / subtracted to the process
visualization.
-10000..+10000 [digit
9
Gain calibration
Percentage value that is multiplied for the process value (allows to
calibrated the working point)
-100.0%..+100.0%, Default: 0.0
ex: to correct the range from 0..1000kg showing 0..1010kg, set the
parameter to -1.0.
10
Calibration
Selection of calibration type for the analog input. After finishing or
deleting calibration proceeding, this parameter will automatically return
to default setting (=Disabled).
Disabled (Default)
Sampling value
], Default: 1000.
1
], Default: 0.
1
1
], Default: 1000.
], Default 0.0.
1
Full-scale % value
mV/V Value
User manual - DP400S -
35