5.1.e
Relay Q2 output
1
2A 230V
Resistive
Q2
1/8HP
2
5.1.f
Output AO1 mA
9
AO1
0/4..20mA
8
5.1.g
Digital Input 1
10
+12 V DC
DI.1
11
(PnP)
Contacts capacity: 2A/250 Vac for resistive loads.
NB: see picture below
Electrical endurance Q1 / Q2.
2 A, 250 Vac, resistive load, 10
20/2 A, 250 Vac, cosφ = 0.3, 10
Pins 8-9: analog output in mA configurable by parameters
as retransmission of process or alarm setpoints (see
parameters 112-116).
PNP digital input.
Digital input configurable by parameter 95
Short-circuit pins 10 and 11 to activate digital input 1.
5
operations.
operations.
5