Modbus
Description
Address
5002
Parameter 1 (H)*
...
...
5300
Parameter 150 (H)*
6001
Data logger: firmware version (H)
6002
Data logger: firmware version (L)
6003
Data logger: sensor type (H)
6004
Data logger: sensor type (L)
6005
Data logger: decimal point (H)
6006
Data logger: decimal point (L)
6007
Data logger: measure unit (H)
6008
Data logger: measure unit (L)
6009
Data logger: sampling time in seconds (H)
6010
Data logger: sampling time in seconds (L)
Data logger: flag end memory. 0 indicates that
memory is still available. 1 indicates memory is
6011
saturated and device restarted to store data from
address 6033/6034 (H)
Data logger: flag end memory. 0 indicates that
memory is still available. 1 indicates memory is
6012
saturated and device restarted to store data from
address 6033/6034 (L)
6033
First analog input value stored (H)
6034
First analog input value stored (L)
6035
Second analog input value stored (H)
6036
Second analog input value stored (L)
...
...
8031
Last analog input value stored (H)
8032
Last analog input value stored (L)
9001
Data logger: firmware version (L)
9002
Data logger: firmware version (H)
Read
Reset value
Write
R/W EEPROM
R/W EEPROM
R/W EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM
RO
EEPROM