Option
Function
CLr ALL
Clear calibration
CAL
points
int
Data log interval
year
Year
date
Date format
m/d date
Month and day
time
Time
oFF
Automatic
shutdown
Snd
Sound options
ºC ºF
Temperature
display
READY
Stabilization lock
1
This does not erase calibration data for an ISFET probe.
4. Push ENTER to keep the changes and go back to setup mode.
Note: To exit and not keep the changes, push CANCEL.

Operation

Configure the Bluetooth
Transfer data and control several instrument functions from a remote location to a PC with the
Bluetooth
wireless connection.
Note: The arrow icon shows on the display when the meter is in communication with the SmartLogger II software
on a PC. The arrow icon flashes when data is transferred. Refer to the SmartLogger II documentation to setup the
wireless connection to a PC.
1. Set the power to on. The meter looks for the Bluetooth
2. On the PC, select the option to find or add a Bluetooth
wants to pair with a found device.
3. When the PC prompts for a pass key or PIN, enter the PIN (default = 1234). The PC shows if the
pair is successful and the meter shows the Bluetooth
14 English
Description
Erases all the calibration points. Make sure to calibrate the meter.
Keeps up to 999 data points in intervals from 1 to 1999 seconds (default =
10) in the data log.
Use the number keys to enter the correct year.
Sets the date to mm/dd/yy or dd/mm/yy format.
Use the number keys to set the correct month and day.
Sets the correct time in a 24-hour format. Use the number keys to set the
correct time.
Sets the shutdown parameters from 1 minute to 000 minutes (continuous
power). The meter beeps 1 minute before shutdown. Make sure to power
cycle for the automatic shutdown to occur. Automatic shutdown is disabled
during: data logging, Bluetooth
port.
Sets sound alerts to on or off. Three sounds tell the user about different
functions:
• One beep: Stabilization occurred with the stabilization lock on.
• Two beeps: An error occurred. The error code shows on the display.
Refer to
Troubleshooting
• Three beeps: Measurement stability in calibration mode, regardless of
the stabilization lock setting.
Sets the temperature unit from ºC to ºF.
Sets the stabilization lock to on or off:
• The "READY" icon shows when an endpoint occurs and the stabilization
lock is set to on.
• After stabilization, the display locks the value and the meter ignores slight
measurement changes in pH, conductivity or TDS.
• Sets the stabilization lock to off during titrations or slight change
detection. The display automatically unlocks after a significant
measurement change is found.
wireless connection
transfers and when connected to the USB
on page 17.
connection with a PC.
device. Next, the PC prompts if the user
icon.
1