8.3 Diagnostic Functions

8.3.1 Basic Diagnostic Codes

Tài li u sưu t m c a Trung Sơn - www.tschem.com.vn - 028.3811.9991 - info@tschem.com.vn
The diagnostic mode permits access to system function information that is useful
primarily when the instrument function is in doubt. Hach service technicians use
the information for precise troubleshooting, speeding repairs, and avoiding
unnecessary service returns.
Access diagnostic information by pressing and holding the
3 seconds. Use the
ARROW
number of interest. Press the
Access the Table 6 diagnostic information by entering the appropriate code:
Code
Display
00
bP on/bP of
01
FS Pr/SL Pr
21
Pr ln
22
*
23
*
24
*
25
xxxxx
26
xxxxx
27
xxxxx
28
xxxxx
29
xxxxx
30
xxxxx
31
xxxxx
32
xxxxx
33
xxxxx
34
xxxxx
35
xxxxx
36
xxxxx
37
xxxxx
38
xxxxx
39
xxxxx
41
xxxxx
45
xxxxx
46
xxxxx
47
xxxxx
48
xxxxx
49
xxxxx
50
xxxxx
51
xxxxx
52
xxxxx
53
xxxxx
70
keys to edit the display to read the diagnostic code
key to display the diagnostic value.
ENTER
Table 6 Diagnostic Codes
Description
Keyboard Beeper On/Off
Fast/Slow Print Device
Printer Test
Display Test
Keyboard Test
Memory Test
90° Detector mV, Gain 1
90° Detector mV, Gain 10
90° Detector mV, Gain 100
90° Detector mV, Gain 1000
Forward Scatter Detector mV, Gain 1
Forward Scatter Detector mV, Gain 10
Forward Scatter Detector mV, Gain 100
Forward Scatter Detector mV, Gain 1000
Transmitted Detector mV, Gain 1
Transmitted Detector mV, Gain 10
Transmitted Detector mV, Gain 100
Transmitted Detector mV, Gain 1000
Transmitted Detector mV, Gain 10,000
Transmitted Detector mV, Gain 100.000
Transmitted Detector mV, Gain 1,000,000
A/D Reference Low mV, Gain 1
A/D Reference Low mV, Gain 10
A/D Reference Low mV, Gain 100
A/D Reference Low mV, Gain 1000
A/D Reference Medium mV, Gain 1
A/D Reference Medium mV, Gain 10
A/D Reference Medium mV, Gain 100
A/D Reference Medium mV, Gain 1,000
A/D Reference High mV, Gain 1
A/D Reference High mV, Gain 10