FLASHING
CODE
8
9
10
11
Meaning
·
Fault on Shark
main module
·
·
Communication
error on the Shark
system
·
·
Unknown error
·
Incompatible
remote
Solution
Check all cables and connectors on
the Shark system.
Replace the main module.
Check all cables and connectors on
the Shark system.
Replace the remote.
Check all cables and U-connectors.
An incorrect remote has been
connected. Ensure that main module
and remote code are concordant.
SERVICE MANUAL
FLASHING CODE
See "Checking the
cables" on page
35
See "Replacing
the Shark
Electronics" on
page 28
See "Checking the
cables" on page
35
See "Replacing
the Shark II
Remote" on page
37
See "Checking the
cables" on page
35
See "Replacing
the Shark II
Remote" on page
37
See "Replacing
the Shark
Electronics" on
page 28