Flashing code:
Flashes six times
Flashes seven times Battery on low
Flashes eight times
Flashes nine or ten
times
Flashes eleven times Drive motors overloaded /
Flashes twelve times Compatibility problems
Significance:
Fault/brake fault on l.h.
motor. Connection loose
or faulty or motor faulty
Battery voltage too high
Faulty data transfer
between the modules
overheated
between the modules
Solution:
·
Check the connecting
plus
·
Charge battery
·
Switch lighting to low
battery voltage
·
Check battery charger
-
·
Switch remote on and
off / wait if necessary
·
Remove incorrect
module
SERVICE MANUAL
Flashing code:
See "Checking
the cables" on
page 35
See "Replacing a
motor" on page
24
See instruction
manual
See instruction
manual for charger
Remove electronic
modules with the
exception of the
main module and
the remote. Install
the modules one
after the other in
order to determine
which one is
causing the
disturbance.
See "Replacing
the Shark
Electronics" on
page 28
-
See "Replacing
the Shark
Electronics" on
page 28