Operating Manual Cardiac Rehabilitation Safety Hardware Software Hygiene Part 2: Hardware, description of device for custo guard (custo cardio concept) Operating characteristics: 1, 3-channel ECG, Cardiac rehabilitation MHW 0010 – DK 1696 Version 001 – 21/09/2017...
Page 2
Internet or from a CD to ensure the Operating Manual is complete. MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
The system is not suitable for electrocardiographic monitoring of patients in accord- ance with DIN EN 60601-2-27, e.g. use in intensive care. MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
Neck strap for case Cable adapter for adhesive electrodes (1 piece) ECG transmitter Electrode belt MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) | page 6...
Page 7
Cable adapter for adhesive electrodes (3 piece) Coding rings for cable adapter and ECG transmitter Electrode belt ECG transmitter MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) | page 7...
Page 8
RF for each custo guard, plus an additional custo com RF (n+1) USB hub to connect several custo com RF (e.g. from LogiLink or MOXA) MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
CAUTION: Charging damp ECG transmitters (perspiration, disinfectants) will lead to the oxidation of the electric contacts in the charger. MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
Fit the blood pressure cuff carefully. The mark or the microphone must rest directly on the brachial artery. MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
Page 13
Adjustment from maximum length to medium length Hold the slider with your right hand, grasp the lower layer of material to the left of the slider and pull it through the slider up to the wide point. Adjustment from medium length to minimum length Hold the slider with your left hand and pull the elastic excess length completely to the right through the slider until the slider is back between the black plastic hook and the wide material point.
We also recommend the use of custo sensive ECG electrodes (latex-free, PVC-free, skin-friendly, permeable to air and moisture). MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
We also recommend the use of custo sensive ECG electrodes (latex-free, PVC-free, skin-friendly, permeable to air and moisture). MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
PA, biocompatible in accordance with EN ISO 10993-1, Oeko-Tex® Standard100 Alternative to custo belt ECG cable adapter, length approx. 300 mm MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) | page 16...
Page 17
USB 2.0 port or network connection PCL-enabled (increases printing speed with the suitable driver) MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) | page 17...
IEC 61000-4-8 COMMENT: U is the alternating supply voltage prior to application of test levels MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) | page 18...
Page 19
Within the frequency range of 150 kHz to 80 MHz, the field strength should be less than 10 V/m. MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
Page 20
These guidelines may not apply in every case. The propagation of electromagnetic variables is influenced by ab- sorptions and reflections of buildings, objects and people. MHW 0010 – DK 1696 | Version 001 – 21/09/2017 | custo med GmbH [2] Cardiac Rehabilitation hardware, custo guard (custo cardio concept) |...
Need help?
Do you have a question about the Custo Guard 1/3 and is the answer not in the manual?
Questions and answers