Siemens SITRANS FS230 Operating Instructions Manual page 18

Ultrasonic flowmeters
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Dual enclosure Reflect mount wiring
Referring to the Hi Precision single enclosure reflect mount figure above, connect the Hi Precision mount reflect Dual
enclosure sensor cables to the transmitter ports as follows:
1. Observing the upstream and downstream orientation
2. Tighten all cable glands to obtain optimum sealing.
3. Refer to transmitter FST030 Operating Instructions to program transmitter.
Dual enclosure direct mount wiring
Connect sensor cables to the DSL and transmitter as follows:
1. Fill connector end with Super Lube prior to connecting.
2. Apply Super Lube to the internal threads at the large end of the thread connector.
3. Observing the upstream and downstream orientation
make connection snug.
4. Connect downstream sensor cable
16
, connect sensor cables to transmitter ports.
, connect upstream sensor cable
to transmitter port (1B) and make connection snug.
to transmitter port (1A) and
SITRANS FS230
A5E38755607-AA, 02/2017