Xylem NiCaVis 705 IQ Operating Manual page 44

Optical sensor for nitrate and carbon sum parameters
Table of Contents

Advertisement

What to do if...
Drifting measured
values
OFL
Display of
Measured values
fluctuating heavily
Measured values
too low or too high
44
Cause
– Cleaning efficiency too low - mea-
surement windows becoming
more and more contaminated
– Failure of the optional com-
pressed air cleaning system
– Substantial change of test sample
(sample matrix)
Cause
– Measuring range exceeded
– Foreign matter in the measuring
gap
– Erroneous user calibration
Cause
– Gas bubbles in front of the mea-
surement windows
Cause
– Gas bubbles in front of the mea-
surement windows
Remedy
– Clean the measurement windows
(section 5.2.4)
– Use other cleaning method
– Select different measurement loca-
tion
– Set longer cleaning duration
– Set more frequent cleaning proce-
dures
– Install optional compressed air
cleaning device
– Clean the measurement windows
(section 5.2.4)
– Check all components of the com-
pressed air cleaning system (com-
pressor, compressed air line,
sensor, rinsing openings)
– Carry out new user calibration
Remedy
– Check process
– Clean the measuring gap (section
5.2.3)
– Repeat user calibration. Make sure
the samples for comparison are rep-
resentative.
Remedy
– Check the position within the pro-
cess and the installation position
Remedy
– Check the installation position of the
sensor
ba75953e03
NiCaVis 705 IQ
05/2017