Xylem YSI SensoLyt 700 IQ H Operating Manual
Xylem YSI SensoLyt 700 IQ H Operating Manual

Xylem YSI SensoLyt 700 IQ H Operating Manual

Hide thumbs Also See for YSI SensoLyt 700 IQ H:

Advertisement

Quick Links

OPERATING MANUAL
ba76206e02 12/2020
®
SensoLyt
700 IQ H
IQ S
N
- pH/ORP SENSOR
ENSOR
ET
FOR CLASS I DIVISION 2 APPLICATIONS

Advertisement

Table of Contents
loading

Summary of Contents for Xylem YSI SensoLyt 700 IQ H

  • Page 1 OPERATING MANUAL ba76206e02 12/2020 ® SensoLyt 700 IQ H IQ S - pH/ORP SENSOR ENSOR FOR CLASS I DIVISION 2 APPLICATIONS...
  • Page 2 ® SensoLyt 700 IQ H For the most recent version of the manual, please visit www.ysi.com. Contact 1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241 800-765-4974 Email: info@ysi.com Internet: www.ysi.com © Copyright 2020 Xylem Inc. ba76206e02 12/2020...
  • Page 3: Table Of Contents

    ® SensoLyt 700 IQ H Contents ® SensoLyt 700 IQ H - Contents Overview ............. 5 How to use this component operating manual .
  • Page 4 ® Contents SensoLyt 700 IQ H What to do if ............28 Technical data .
  • Page 5: Overview

    ® SensoLyt 700 IQ H Overview Overview How to use this component operating manual Structure of the IQ Sensor Net Operating Manual IQ S operating ENSOR manual System Operating Manual (Ring Binder) IQ Sensor MIQ Module MIQ Terminal Operating Operating Operating Manual Manual...
  • Page 6: 700 Iq H

    ® Overview SensoLyt 700 IQ H ® Structure of the SensoLyt 700 IQ H ® Fig. 1-2 Structure of the pH/ORP sensor (Example SensoLyt 700 IQ H) Protective hood Temperature sensor Combination electrode (not contained in the scope of delivery) Electrode receptacle Sensor shaft The pH combination electrodes that can be used are avail-...
  • Page 7: Safety

    ® SensoLyt 700 IQ H Safety Safety Safety information 2.1.1 Hazard warnings in this operating manual This operating manual provides important information on the safe oper- ation of the product. Read this operating manual thoroughly and make yourself familiar with the product before putting it into operation or working with it.
  • Page 8: Safe Operation

    ® Safety SensoLyt 700 IQ H tem:  Operating manuals of other components of the measuring system (Power supply units, controllers, accessories)  Safety datasheets of calibration and maintenance equipment (e.g. cleaning solutions). Safe operation 2.2.1 Authorized use ® The authorized use of the SensoLyt 700 IQ H consists of its use as a sensor in the IQ S .Only the operation and running of the...
  • Page 9: Hazardous Location Use

    ® SensoLyt 700 IQ H Safety Hazardous location use GEFAHR Explosion hazard. Only hazardous location rated sensor models ("Hazloc Sensor") must be used in hazardous locations. Read the name plate on the sensor shaft and verify that the sensor is rated for your specific application.
  • Page 10: Commissioning

    ® Commissioning SensoLyt 700 IQ H Commissioning Scope of delivery ®  SensoLyt 700 IQ H  The sensor is equipped with a protective hood and protective caps  Operating manual. Installation Safety guidelines for installation in a GEFAHR hazardous location Explosion hazard.
  • Page 11 ® SensoLyt 700 IQ H Commissioning them dry using compressed air). Do not suspend the sensor on the sensor connection cable. Use an armature or electrode holder. NOTE ® The pH/ORP sensor SensoLyt 700 IQ H may only be submersed in conjunction with a mounted combination electrode.
  • Page 12: Commissioning / Getting The Sensor Ready For Measuring

    ® Commissioning SensoLyt 700 IQ H SACIQ Fig. 3-1 Connecting the sensor Commissioning / Getting the sensor ready for mea- suring A KCI-filled plastic cap is mounted on the tip of the combi- nation electrode to keep the electrode active during storage (or during longer pauses in measuring).
  • Page 13 ® SensoLyt 700 IQ H Commissioning Screw the protective cap off the plug head connector of the electrode. Screw the electrode into the plug head socket of the sensor. Push the unit into the sensor up to the stop. ba76206e02 12/2020...
  • Page 14: 700 Iq H

    ® Commissioning SensoLyt 700 IQ H ACHTUNG Push the connected electrode into the sensor right up to the stop so that the connection is watertight. Leakiness can cause the sensor to be destroyed. For measuring, pull the KCI-filled plastic cap off the combina- tion electrode.
  • Page 15 ® SensoLyt 700 IQ H Commissioning Setting Selection/values Explanation  mV Unit of the measured value in the measured Measuring mode value display.  pH  °C Unit of the temperature measured value Temperature mode (Celsius, Fahrenheit).  °F  CAL TEC AUTO ...
  • Page 16 ® Commissioning SensoLyt 700 IQ H Setting Selection/values Explanation The temperature compensation enables the Temp. adjustment -1.5 K ... +1.5 K adjusting of the temperature sensor to a ref- erence temperature measurement (offset- ting the zero point by ±1.5 K). Notes: ...
  • Page 17: Measurement / Operation

    ® SensoLyt 700 IQ H Measurement / operation Measurement / operation Measuring WARNUNG Chemical or biological hazard. Contact with the sample can be harmful to the user. Depending on the type of sample, suitable protective measures must be taken (protective clothing, gog- gles, etc.).
  • Page 18: Calibrating With Cal Tec Auto

    ® Measurement / operation SensoLyt 700 IQ H Maintenance condition During the calibration procedure the sensor is in the so-called maintenance condition. All linked outputs remain in their current status. After the calibration procedure has been finished the maintenance condition has to be switched off manually. For detailed information on the maintenance condition please refer to the relevant IQ S ENSOR system operating manual.
  • Page 19: Calibrating With Cal Con 2P

    ® SensoLyt 700 IQ H Measurement / operation Display Explanation Follow the instructions on the dis- * Rinse the sensor. play. As soon as the measured * Immerse the sensor in the first value is stable, the next display buffer solution. indication appears.
  • Page 20: Calibrating With Cal Con 1P

    ® Measurement / operation SensoLyt 700 IQ H Display Explanation Follow the instructions on the * Rinse the sensor. display. As soon as the mea- * Immerse the sensor in the sec- sured value is stable, the next ond buffer solution. display indication appears.
  • Page 21: Calibration Result

    ® SensoLyt 700 IQ H Measurement / operation Display Explanation The values determined for Slope Successfully calibrated. and Asymmetry potential are dis- End of the CAL_CON_1P calibra- played. tion. Calibration is finished. Confirm with <OK>. The display returns to the mea- sured value display.
  • Page 22: Calibration History

    ® Measurement / operation SensoLyt 700 IQ H 4.2.6 Calibration history Calibration history Currently active calibration Chronological list of the last calibration procedures ® Fig. 4-1 Calibration historySensoLyt 700 IQ H The calibration history provides the following information:  Date of the calibration ...
  • Page 23: Maintenance And Changing The Electrode

    ® SensoLyt 700 IQ H Maintenance and changing the electrode Maintenance and changing the elec- trode General maintenance instructions ® The SensoLyt 700 IQ H pH/ORP sensor operates maintenance-free. Please read the maintenance of the combination electrode in the relevant operating manual. WARNUNG Chemical or biological hazard.
  • Page 24: Replacing The Electrode

    ® Maintenance and changing the electrode SensoLyt 700 IQ H tion cables. Replacing the electrode If it is necessary to replace an electrode, proceed as follows: Unscrew the protective hood from the sensor. Use the protective hood as a tool to lever out the electrode. Carefully pull out the electrode until the plug head screwed fit- ting can be seen.
  • Page 25 ® SensoLyt 700 IQ H Maintenance and changing the electrode Screw in a new combination electrode. Push the unit into the sensor up to the stop. For measuring, pull the KCI-filled plastic cap off the combina- tion electrode. ba76206e02 12/2020...
  • Page 26: Disposal

    ® Maintenance and changing the electrode SensoLyt 700 IQ H Screw the protective hood onto the sensor. Calibrate the sensor and the electrode with the measuring sys- tem (see section 4.2 C ALIBRATION Disposal Sensor We recommend disposing of the sensor as electronic refuse. Combination electrodes If no official regulations apply to the contrary, used and defective elec- trodes can be treated as household waste.
  • Page 27: Replacement Parts And Accessories

    ® SensoLyt 700 IQ H Replacement parts and accessories Replacement parts and accessories Combination electrodes pH combination Model Order no. electrodes ® 109 115Y SensoLyt ® 109 119Y SensoLyt ® 109 117Y SensoLyt ® 109 118Y SensoLyt SEA-HP ORP combination ®...
  • Page 28: What To Do If

    ® What to do if ... SensoLyt 700 IQ H What to do if ... No measured value Cause Remedy Connect the sensor Sensor not connected Refer to log book Unknown Measurement does not Cause Remedy function Pull off watering cap and calibrate Watering cap still on the elec- trode Connect the electrode...
  • Page 29 ® SensoLyt 700 IQ H What to do if ... Cause Remedy Correct the instrument setting Incorrect instrument setting (Measuring mode pH or mV) Measured value flashing Cause Remedy Maintenance condition is active – If the maintenance condition was activated manually (e.g. by pressing the <C>...
  • Page 30: Technical Data

    ® Technical data SensoLyt 700 IQ H Technical data Measuring characteristics Measuring principle Potentiometric measurement using a combination electrode; Integrated microprocessor electronics, shielded 2-wire connection for power and data transmission. Measuring range 0.00 ... 14.00 pH (depending on the electrode) -2000 mV ...
  • Page 31 ® SensoLyt 700 IQ H Technical data 6 x10 Pa (6 bar) * Max. allowed overpressure with electrode ® SensoLyt ECA installed Pa (10 bar) ** Max. allowed overpressure with electrode ® SensoLyt SEA-HP installed * temperature dependent (see safety instruction below) ** in the entire temperature range ®...
  • Page 32: General Data

    ® Technical data SensoLyt 700 IQ H Operating position Fields of application ® Stationary measurements in wa- SensoLyt 700 IQ H ter/wastewater applications General data Dimensions (mm) 40.0 SACIQ socket... Weight with electrode approx. 970 g (without sensor connection cable) ®...
  • Page 33: Instrument Safety

    ® SensoLyt 700 IQ H Technical data Instrument safety 8.4.1 General instrument safety Applicable norms – UL 61010-1 – CAN/CSA C22.2 # 61010.1 8.4.2 Hazardous location ratings Applicable directives In addition to the standards listed in chapter 8.4.1, the ® and standards SensoLyt 700 IQ H conforms to the following directives and stan-...
  • Page 34: Indexes

    ® Indexes SensoLyt 700 IQ H Indexes Explanation of the messages This chapter contains a list of all the message codes and related ® message texts for the SensoLyt 700 IQ H sensor. Information on  the contents and structure of the log book and ...
  • Page 35: Informative Messages

    ® SensoLyt 700 IQ H Indexes Message code Message text EI4311 Operational voltage too low, no operation possible * Check installation and cable lengths, Follow installation instructions * Power supply module overloaded * Check terminal and module connections * Defective component, replace components ES1311 Component hardware defective...
  • Page 36: Status Info

    ® Indexes SensoLyt 700 IQ H Status info The status info is a coded piece of information on the current status of a sensor. Each sensor sends this status info to the controller of the IQ S . The status info of sensors consists of 32 bits, each of ENSOR which can have the value 0 or 1.
  • Page 37: Contact Information

    ® SensoLyt 700 IQ H Contact Information Contact Information 10.1 Ordering & Technical Support Telephone: (800) 897-4151 (937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET Fax: (937) 767-1058 Email: info@ysi.com Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 Internet: www.ysi.com When placing an order please have the following information available:...
  • Page 38 ® Contact Information SensoLyt 700 IQ H ba76206e02 12/2020...
  • Page 40 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation. For more information on how Xylem can help you, go to www.xyleminc.com 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241;...

Table of Contents