Xylem NiCaVis 705 IQ Operating Manual page 35

Optical sensor for nitrate and carbon sum parameters
Table of Contents

Advertisement

NiCaVis 705 IQ
Result
Sensor
after
check (H2O dest.)
Result
Zero
after
adjustment
ba75953e03
05/2017
15.
Confirm each checklist with <OK> until the
appears.
Up to this point, you can break off the calibration procedure at any
time with the <ESC> key. The system continues to work with the old
calibration data. The maintenance condition, however, has to be
switched off again in any case.
After starting the calibration with the <OK> key (step 16.) you can no
longer cancel the calibration procedure.
16.
Confirm with <OK>.
The measurement is started. After the end of the measurement the
result appears. With the
calibration history at the same time.
Rating number
Meaning
Sensor check (H2O dest.)
Zero range
(-40 ... +40)
(0 = no deviation)
Positive number
Measured values were too low.
(+40 ... +1000)
Possible cause:
The measurement windows are contaminated.
Negative num-
Measured values were too high.
ber
Possible cause:
(-40 ...-1000)
During the last zero adjustment, the measurement
windows were contaminated more than during the
current sensor check.
If the rating number is not in the zero range:
Repeat the cleaning procedure and the
number no longer changes significantly.
If the rating number is still outside the zero range, a new zero adjustment is nec-
essary.
Result
Meaning
OK
Zero adjustment
Error
Zero adjustment
ful.
If the result is erroneous, clean the sensor and all equipment once again. Then
Zero adjustment
carry out a new
are absolutely clean.
Start measurement
Zero adjustment
, the result is entered in the
successful (OK)
Sensor check (H2O dest.)
Sensor released for
measurement
successful.
yes
not success-
No ("----")
and when doing so make sure the conditions
Measurement / Operation
display
until the rating
35