Chamber Temperature Setting - Sanyo MPR-1411 Instruction Manual

Pharmaceutical refrigerator
Hide thumbs Also See for MPR-1411:
Table of Contents

Advertisement

CHAMBER TEMPERATURE SETTING

Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
o
temperature is 10
C.
Note: The unit is set at the factory with a chamber temperature of 5
Table 1 Basic operation sequence (Example: Chamber set temperature 10
Description of operation
1
Connect the power supply plug.
2
Press set key.
Set to 010 with the numerical value
3
shift key and digit shift key.
4
Press set key.
Note:
• The temperature set mode returns to the temperature display mode automatically when 90 seconds has
passed without any key operation. In this case, the chamber temperature setting is not accepted.
• The available set range of chamber temperature is between 2 and 23
occurred if the set temperature is lower than 3
Key operated
The current chamber temperature is
----
displayed.
The second digit is flashed.
SET
By pressing key, the settable digits
moves.
When pressed, the figure of settable
digit increases.
Set temperature is memorized and the
SET
current chamber temperature is
displayed.
o
C.
- 54 -
o
C.
o
C)
Indication after operation
o
C. The partial freezing may
16