Setting Of Alarm Resume Time - Sanyo MPR-1411 Instruction Manual

Pharmaceutical refrigerator
Hide thumbs Also See for MPR-1411:
Table of Contents

Advertisement

SETTING OF ALARM RESUME TIME

The alarm buzzer and remote alarm are silenced by pressing he alarm buzzer stop key (BUZZER) on the
control panel during alarm condition. The buzzer and remote alarm will be activated again after certain
suspension if the alarm condition is continued.
procedure shown in the Table 5 below.
The example in the table is based on the assumption that the desired duration is 20 minutes.
Note: The duration is set in 30 minutes at the factory.
Table 5
Procedure for setting of alarm resume time (Example: change from 30 minutes to 20
minutes)
Operation
1
Press and hold the numerical value
2
shift key for about 5 seconds.
Set to F25 by using digit shift key
3
and numerical value shift key.
4 Press the set key.
Set to 020 by using numerical value
5
shift key.
6 Press the set key.
Note:
• The settable alarm resume times are 10, 20, 30, 40, 50, or 60 minutes (The set values are 010, 020,
030, 040, 050, and 060). The buzzer would not reset if the resume time is set in 000.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
19
The suspension time can be set by following the
Key operated
The current chamber temperature
-----
is displayed.
The first digit of the temperature
display blinks.
Pressing the key leads the change of flashed
digit.
Pressing the key shifts up the figure
of the current digit.
The current set value is displayed and
the second digit of the temperature
SET
display flashes.
Pressing the key shifts up the figure in
the second digit.
The value is stored in memory and
SET
the current chamber temperature is
displayed.
- 57 -
Indication after operation

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mpr-1411r

Table of Contents