REFRIGERATOR COMPONENTS

Control panel

1
2
3
7
6
5
4
1. Door check indicator (DOOR): The red LED lamp is lit when the outer door is opened.
2. Alarm lamp (ALARM): This lamp is flashed when the audible alarm is activated.
3. Digital temperature indicator:
This indicator shows the present chamber temperature or set
temperature.
4. Numerical value shift key (
): Pressing this key in the setting mode causes the numerical value to
shift. ON-OFF of key lock can be selected by pressing this key in the key lock mode.
5. Digit shift key (
): Pressing this key in the setting mode causes the changeable digit to shift. Key
lock is available by pressing this key for more than 5 seconds in the temperature display mode. Refer to
page 20 for the key lock.
6. Set key (SET): Temperature setting mode is led by pressing this key. Once the key is pressed, the
changeable digit is flashed. Pressing this key again after setting desired temperature, the setting is
stored into computer memory. If there is no key operation for 90 seconds during the temperature setting
mode, the temperature setting mode is invalid automatically. See page 16 for the details.
7. Buzzer stop key (BUZZER): To silence the audible alarm, press this key. The remote alarm is also
silenced by pressing this key. (The buzzer cannot be stopped during remote alarm testing.)
- 48 -