Cable Connections - 6-Core Sensor Cable / Endura Az25 Remote Transmitter To Endura Az25 Probe - ABB Endura AZ25 Series User Manual

Combustion oxygen monitor, probe
Hide thumbs Also See for Endura AZ25 Series:
Table of Contents

Advertisement

Endura AZ25 series probe
Combustion oxygen monitor
4.2.2 Cable connections – 6-core sensor cable / Endura AZ25 remote transmitter to Endura AZ25 probe
Transmitter
terminal block color
Green
White
Red
Black
Orange
Blue
SCN (not used) – see Fig. 4.2
sensor cable screen (drain) is
connected at internal earth stud only
Table 4.4 Cable connections – 6-core sensor cable / Endura AZ25 remote transmitter to Endura AZ25 probe
Factory-made
connections
(at probe inner
terminal connectors)
*Cut the screen, screen foil and drain at probe head back and isolate only – do not connect these items at probe
**Connect drain from 6-core cable at transmitter to the transmitter internal earth stud only –
do not connect to transmitter terminal SCN connector or transmitter external earth
Fig. 4.2 Cable connections – 6-core sensor cable / Endura AZ25 remote transmitter to Endura AZ25 probe
22
6-core screened remote transmitter
to probe cable – wire colors
Screen (drain) – see Fig. 4.2 for sensor
cable screen connection requirements
Endura AZ25 probe
User-made connections
– 6-core sensor cable to
Endura AZ25 AutoCal
unit
Screen / drain of 6-core sensor
cable at Endura AZ25 probe cut
back and isolated*
Endura AZ25
remote transmitter
Green
White
Red
Black
Orange
Blue
(no screen connection made at probe head)
Terminal connections at Endura AZ25 probe
T/C +
T/C –
Cell +
Cell –
ACJC
ACJC
Terminal connections at
Endura AZ25 remote transmitter
Cell – (Black)
ACJC (Orange)
ACJC (Blue)
T/C + (Green)
T/C – (White)
NOT USED
SCN (Screen)
Connect drain at Endura AZ25
transmitter end to transmitter
internal earth stud only**
4 Electrical connections
Probe / remote transmitter
connection
TC +
TC –
Cell +
Cell –
ACJC
ACJC
Screen / drain – see Fig. 4.2 for sensor
cable screen connection requirements
at transmitter
Green
User-made
White
connections –
Red
6-core sensor
Black
cable to
transmitter
Orange
Blue
6-Core sensor
cable wire colors
Cell + (Red)
OI/AZ25P–EN Rev. E