Endura Az25 System Without Autocal; Cable Preparation - 6-Core Sensor Cable / Endura Az25 Remote Transmitter To Endura Az25 Probe - ABB Endura AZ25 Series User Manual

Combustion oxygen monitor, probe
Hide thumbs Also See for Endura AZ25 Series:
Table of Contents

Advertisement

Endura AZ25 series probe
Combustion oxygen monitor

4.2 Endura AZ25 system without AutoCal

4.2.1 Cable preparation – 6-core sensor cable / Endura AZ25 remote transmitter to Endura AZ25 probe
Endura AZ25 remote transmitter
Transmitter
end – cut to
10 mm (0.4 in.)
Cut outer insulation
sheath and screen foil
back – leave 110 mm
(4.4 in.) drain wire exposed
and sleeve to 100 mm (4 in.)
Fig. 4.1 Cable preparation – 6-core sensor cable / Endura AZ25 remote transmitter to Endura AZ25 probe
OI/AZ25P–EN Rev. E
Green (TC+)
Screen (with drain)
Blue (ACJC)
Orange (ACJC)
Black (Cell –)
Wire colors – 6-core sensor cable
6-core screened sensor cable, probe to transmitter,
maximum length 100 m (328 ft.) – coded option
Transmitter end –
cut to 100 mm (4 in.):
including drain wire
Endura AZ25 probe
White (TC–)
Red (Cell +)
Probe end –
cut to 150 mm (6 in.)
Cut the outer insulation
sheath, screen foil and
drain wire back and isolate
all three items
4 Electrical connections
Probe end –
cut to 10 mm
(0.4 in.)
21