To Purge The Ci Calibration Valve - HP 5973 MSD Hardware Manual

Table of Contents

Advertisement

To purge the CI calibration valve

After adding new PFDTD to the calibrant vial, you must purge the air out of
the vial and valve.
1 If the vacuum gauge controller is on, turn it off.
2 Turn on
3 Turn on
4 In Diagnostics and Vacuum Control view, select
Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
C A U T I O N
After removing the calibrant vial, you
will result in severe contamination of the ion source and damage to the electron multiplier.
Gas A.
Purge.
7R SXUJH WKH &, FDOLEUDWLRQ YDOYH
Purge Cal Valve
purge the calibration valve. Failure to do so
must
 Maintenance
under the
99